8.8 General items
HVA4321
8-8
8.8 General items
Input/output ground
All signal input/output (BNC connector) excluding the terminal block for ON/OFF status signal
contact output, and the total current terminal of the high voltage output terminal (OUTPUT Lo) are
connected to the casing.
Power supply
Range of power supply voltage
Single phase AC 100 V ±10 %
Frequency of power supply
50 Hz/60 Hz ±2 Hz
Power supply fuse
AC 250V, 10 A, Time lag
φ
2 x 20 mm
Power consumption
800 VA or less
Installation environment of unit
Installation position of unit
Horizontal (10° or less)
Cooling method of the unit
Forced air cooling
(Intake from back, exhaust from sides)
Sufficient space is provided at the back and sides of the
unit not to disturb air cooling function.
Environment condition
Performance guarantee
Temperature
+5
℃~
+35
℃
Humidity
10
~
85 %RH (No condensation)
Operation guarantee
Temperature
0
℃~
+40
℃
Humidity
10
~
85 %RH (No condensation)
Storage Temperature
-10
℃~
+50
℃
Humidity
10
~
80 %RH (No condensation)
Insulation resistance
Power supply input to casing
30 M
Ω
or more (with DC 500 V)
Voltage resistance
Power supply input to casing
AC 1500 Vrms for 1 minute
External dimensions
440 mm (width) x 177 mm (height) x 450 mm (depth) (excluding handle and protrusions)
Weight
Approx. 18 kg
Summary of Contents for HVA4321
Page 1: ...INSTRUCTION MANUAL NF Corporation HVA4321 10kV AC DC AMPLIFIER HVA4321 10kV AC DC AMPLIFIER...
Page 2: ......
Page 4: ......
Page 8: ...HVA4321 iv...
Page 16: ...Figures HVA4321 xii...
Page 18: ...HVA4321 1 2...
Page 24: ......
Page 30: ...2 3 Grounding and power supply connection HVA4321 2 6...
Page 32: ......
Page 50: ......
Page 52: ......
Page 54: ......
Page 56: ......
Page 116: ......
Page 127: ...6 Troubleshooting 6 1 When suspecting a fault 6 1...
Page 128: ......
Page 132: ......
Page 138: ......
Page 148: ......
Page 150: ......
Page 151: ......