5.1 Measuring method
HVA4321
5-1
5.1 Measuring
method
Measuring method of output voltage
Measure the output voltage with the high voltage probe (P6015A by Tektronix). To measure frequency
characteristics, compensate the frequency characteristics of the
high voltage probe using equalizing function of the frequency characteristic analyzer (FRA5095 by
NF Circuit Design Block).
Figure5-1. Compensation of probe
The measuring method of frequency characteristics for output voltage and voltage monitor is as shown
in the figure below.
Connect load as required. Take measurements with the load cable length of approx. 15 cm for
resistance load and with the load cable length of approx. 5 cm for capacity load.
Figure5-2. Measuring method of output voltage
信 号 出 力
ch1
入 力
周波数 特性分 析器
高 圧 プ ロ ー ブ
10Vr ms
出 力
Frequency characteristic analyzer
Signal output
10 Vms output
Ch 1 input
High voltage probe
Frequency range:
1 Hz to 2 MHz
After measurement, store the measurement
results in the EQL memory.
信 号 入 力
電 圧 モ ニ タ
出 力電 圧
高 電 圧 プ ロー ブ
HVA4321
周 波 数 特 性 分 析 器
Frequency characteristic analyzer
Signal input
Voltage monitor
Output voltage
High voltage probe
Summary of Contents for HVA4321
Page 1: ...INSTRUCTION MANUAL NF Corporation HVA4321 10kV AC DC AMPLIFIER HVA4321 10kV AC DC AMPLIFIER...
Page 2: ......
Page 4: ......
Page 8: ...HVA4321 iv...
Page 16: ...Figures HVA4321 xii...
Page 18: ...HVA4321 1 2...
Page 24: ......
Page 30: ...2 3 Grounding and power supply connection HVA4321 2 6...
Page 32: ......
Page 50: ......
Page 52: ......
Page 54: ......
Page 56: ......
Page 116: ......
Page 127: ...6 Troubleshooting 6 1 When suspecting a fault 6 1...
Page 128: ......
Page 132: ......
Page 138: ......
Page 148: ......
Page 150: ......
Page 151: ......