7.4 Performance test
HVA4321
7-3
7.4 Performance
test
This section describes a simple performance test of the unit.
7.4.1
Test of constant voltage mode
The gain accuracy and the monitor accuracy when the standard signal input is given in constant voltage
mode are checked.
Give voltage input from the reference DC power supply to the external DC bias signal input. Make sure
that the output is +10 kV for +10V input.
Make sure that the error between the actual output and the value of the output monitor is 0.3% or less.
Figure7-2. Check method of constant voltage mode
HVA4321
直流基準電源
電圧計( 出力モニタ )
電圧計( 出力)
分 圧 器
DC standard power supply
Voltage
divider
Voltmeter
(output monitor)
Voltmeter
(output)