9
Pause Mode
Interrupting a Nexalin ADI Therapy session is not recommended and should only be performed by a
NCT. However, if required, it is possible to pause the therapy session. While in Pause Mode, the
Nexalin ADI Device will continually beep and the CURRENT LEVEL and TIME windows will flash.
1. To pause a Nexalin ADI Therapy session, press the CURRENT OFF button at any time during
the treatment.
2. The current output will ramp down to zero and turn off. However, the Nexalin ADI Device
will remain on until the POWER SWITCH is turned off.
3. The CURRENT LEVEL and remaining session TIME will flash on the front panel, and an
audible warning tone will continue to sound indicating the session has been paused.
NOTE: In order to stop the Nexalin ADI Device from flashing and beeping, press the
CURRENT OFF button again.
4. Resume the session by pressing the CURRENT ON button. The CURRENT LEVEL will ramp up
from zero to 4.00 over a period of approximately 20 to 30 seconds.
5. The session TIME will resume counting down and the therapy session will continue.
Default Nexalin ADI Device Parameters
1. The default CURRENT LEVEL for the Nexalin ADI Device is 4.00 mA. This is not adjustable.
2. The default therapy session TIME for the Nexalin ADI Device is 40:00 minutes. This is not
adjustable.
CAUTION!
To ensure the safety and effectiveness of the therapy, use only Nexalin Electrodes approved by the
manufacturer.
Failure to discharge static electricity prior to handling the patient cable may cause damage to the
device.
Never power the Nexalin ADI Device ON or OFF with a patient connected.
The patient should avoid unnecessary movement while connected to the Nexalin ADI System as the
electrode clips may become disconnected.
Place the electrode cables away from traffic areas.
Do not bend electrode cable and lead wires as they do not withstand rough handling.
If the diagnostic self-test does not complete successfully or an LED remains lit, do not use the device.
Contact the Nexalin Support Center for assistance.