13
All Other Error Codes
If an error code other than 9 is displayed in the TIME window, it indicates an internal fault has been
detected and the device will stop functioning. Perform the following steps to attempt to correct the
error:
1. Ensure the patient is disconnected from the Nexalin ADI Device.
2. Power off the Nexalin ADI Device using the ON/OFF switch.
3. Verify the Nexalin ADI Device is properly plugged into a functioning power source.
4. Wait approximately ten seconds, then power the device on.
5. Upon powering on, the Nexalin ADI Device will perform the diagnostic self-test.
6. The powering on process has successfully completed when the CURRENT LEVEL window reads 4.00
and the session TIME window reads 40:00.
If the error code persists, the device must be returned to the manufacturer for replacement. Contact
the Nexalin Support Center for instructions on returning the device.
CAUTION!
Never power the Nexalin ADI Device on or off with a patient connected.