9. ERROR MESSAGES
Code
Message
Description/Solution
USER COMMUNICATIONS
The shielded door connected to the
X-ray device is open.
Close the door or any other device which
ensures a safe X-ray emission for the
operator.
WARNING
0.1
X-ray
control
released
during
exposure.
Keep X-ray emission control pressed until
the end of the procedure.
0.2
X-ray control not released at the end
of the examination.
1) X-ray control pressed for too long after
the end of the exposure (>15s).
2) Reset command sent while the X-ray
control was pressed.
3) The device exited stand-by mode while
the X-ray control was pressed.
0.3
Emergency stop button onboard the
machine pressed.
Release the emergency stop button.
0.13
2D detector start-up failure
Check contacts of removable sensors.
Disconnect the sensor and connect it
again.
0.16
Move 2D Detector to the PAN
position.
Move 2D Detector to the PAN position.
0.17
Move 2D Detector to the CEPH
position.
Move 2D Detector to the CEPH position.
0.18
Check
the
Craniostat
correct
position.
Check the Craniostat correct position.
0.22
Open
door
detected
(active
interlock).
Check interlock switch (unclosed door?).
0.23
Image not downloaded from the
equipment, download it on PC or
USB before proceeding.
New exposure required, while the previous
one has not yet been downloaded.
0.24
Wait for confirmation before pressing. X-ray control pressed before starting the
cbct examination procedure. For CBCT
examinations, launch “Patient scanning”
from NNT and wait for the explicit request
before pressing the X-ray control.
0.25
CEPH enabled.
Only upon the first start-up after CEPH
arm installation, contact the technical
service department if the problem persists.
0.27
Device not configured.
Configure it from PC.
0.28
Device not calibrated.
Calibrate the unit.
0.29
Make the patient release the bite and
press reset to move the chin rest to a
new position.
The chin rest performs a long movement,
therefore the patient must release the bite
and allow the support movement.
0.30
Key pressed upon start-up.
Keys pressed upon start-up detected.
Check that all keyboard keys are working
properly.
For any other error, turn off the unit, wait for 30 sec and turn it on again. If the problem persists, please contact the
technical service department.