6.2.3. EXAMINATION SETTINGS
After selecting the examination category and type and the
anatomical regions concerned, the SETTINGS screen is
shown.
This screen can be displayed also by selecting a
“favourite” examination in the Home page.
This screen includes:
1 - Summary of FOV and position chosen:
touch this area to go back to the ANATOMIES screen,
where the choice made can be modified.
2 - DOSE profile:
in this area, it is possible to select the dose mode of
administration. Technical features are set to ensure the
best possible quality of the final image, minimising the
radiation dose to the patient according to the chosen
profile:
LOW DOSE (ECO): scanning time and X-ray parameters
are reduced to the minimum necessary for obtaining
standard-resolution images
REGULAR scanning time and X-ray parameters are those
optimal for obtaining standard-resolution images
BEST QUALITY: scanning time and X-ray parameters are
those which allow obtaining the best image resolution
Depending on the choices made, the estimated dose indication will be updated (DAP, kV and mAs), according to the
FOV size and the dose profile chosen.