![Newtech 6300A Operator'S Manual Download Page 5](http://html1.mh-extra.com/html/newtech/6300a/6300a_operators-manual_1680810005.webp)
6300A Operator’s Manual
- 1 - NewTech Medical Limited
1 GENERAL
6300A
microprocessor ventilator is pneumatic driven, electronic-controlled, pressure-limit and time-
cycled multifunctional ventilator also with perfect multi-ventilating mode and monitoring system. Its
panel configuration refers to the figure.
1.1 Main
Technical
Specification
1.1.1
The performance of Airway System
a) Safety pressure of the airway system: no more than 6kPa;
b) Inspiration Fi - oxygen: no more than 45%
1.1.2 Ventilation
mode
a) A/C
b) A/C+ SIGH
c) SIMV (f / 2, f / 4)
d) SPONT
1.1.3 Ventilation
Performance
a) Respiration frequency f
6
~
60 times/min
b) I: E Ratio: 1:4, 1:3, 1:2, 1:1.5, 1:1,2:1,
±
15%
c) Tidal volume 0, 80~1200mL,
±
15%
d) Ventilating volume: >18L/min
1.1.4 Monitor
Performance
a) Tidal volume
digital display with error of
±
20%
b) Ventilation frequency
digital display
c) Airway pressure analogue display
1.1.5 Alarm
Performance
a) Airway pressure high limit alarm
2~ 6kPa
±
15%
b) Airway pressure low limit alarm
0~0.5kPa
±
100Pa; 0.5 ~ 2kPa
±
15%
c)
Power-off
alarm
>120s
d)
Silence
<120s
1.1.6
Performance of the whole set
a) Power <50W
b) Noise of whole set <65dB(A)
c) Electrical safety Meet requirements for I type B mode equipment specified in IEC60601-1
<<Medical Electrical equipment: Part one: General requirement for safety>>