Newtech 6300A Operator'S Manual Download Page 21

6300A Operator’s Manual   

 

                                         

      - 17 -                                                              NewTech Medical Limited 

 

 

6.2 Maintenance 

 

6.2.1 Maintainability 

The 

6300A

 ventilator is designed to place reliability at the principal position, and it is required that 

the mean time of break free (MTBF) of the whole machine be greater than 2000 hours. But human 
operates the machine; the operating life of the machine in a great extent is dependent on the 
human factors. Correct operation and careful maintenance will greatly prolong the life of the 
machine, and reduce the appearance of faults. The repairing operation is divided into maintenance 
and repair. By maintenance it is meant the fault location and removing after fault appears. The 
repair is under the charge of the repairing personnel from the factory or specified maintenance 
point; the user should not carry out repairing by him. For the doctor or nurse who operates the 
ventilator, stress should be laid on the regular maintenance of the machine. 
 

6.2.2  Disinfect, cleaning, drying and correctly install of the easy 

damageable part. 

The pipeline for connecting the machine and patient is easy to be polluted, so disinfectant is 
needed, and the method is shown in the operation manual. The airway channel

connector and 

sensor, etc. After disinfectant should be correctly installed in position without gas leakage. After 
installation, trial running should be conducted, and the machine can be connected to patient and 
used only after it is operated normally. It should be specially specified that the arrowed head on the 
flow sensor means the flow direction, and it should not be installed in reverse. Also the sensor of 
the inspiration circuit and that of the expiration circuit can’t be mistaken. Generally, the expiration 
channel is easy to be polluted, thus just disinfecting the expiratory flow sensor is OK. After 
disinfectant and cleaning, it must be dried, otherwise, if there is water on the wire mesh, the 
resistance will increase, the expiration flow will become inaccurate, thus the reading on the expired 
minute volume meter will be affected. 
Those parts that directly touch the patient's body are easily polluted, the sterilization must be done 
after every use as explained before. It must be here emphasized is that: One must do strictly the 
correct installation after every cleaning and sterilization. The recovered machine must be tested 
and ensured in good condition before putting in reuse for patients.   
 

6.2.3 

Maintaining good contact for connectors 

The connector comprises airway connector and circuit connector. For all the joints, rubber 
connectors and delivering pipes of the airway, good contact should be ensured so as to guarantee 
the gas circulating without gas leakage. For all the connectors in the circuit, check should be 
conducted before operation, especially connectors of inspiration valve and expiration valve, as they 
have no locking device, main check should be carried out to prevent them from loosening. 

 

6.2.4 

Correctly operation of panel knob 

The panel knob should be operated carefully with slight motion. In special, the potentiometer knobs 
should not be screwed with force when the terminal position is attained as this may lead to 
dislocation. After dislocation, the set value is not accurate, which even may cause abnormal 
operation of the machine. 

Summary of Contents for 6300A

Page 1: ...L VENTILATOR 6300A NewTech Medical Limited 8400 Normandale Lake Boulevard Suite 920 Bloomington 55437 Minnesota U S A Tel 1 952 837 2688 Fax 1 952 400 8947 http www newtech medical com E mail info new...

Page 2: ...LATION AND CHECK 8 3 1 Ventilator Component 8 3 2 Ventilator Installation 9 3 3 Check 9 4 OPERATION 11 4 1 Parameters setting 11 4 2 Operation 12 5 OPERATION METHODS OF VARIOUS VENTILATION MODES 13 5...

Page 3: ...m the machine Always use a fresh bacterial filter and test circuit A hazard to the health of the service person may result z Hazardous voltages exist within this medical device Always observes appropr...

Page 4: ...before putting the machine into operation and strictly follow all the procedures of operation Our company guarantees the quality of the product Any time and in any case when you need our help we are...

Page 5: ...PONT 1 1 3 Ventilation Performance a Respiration frequency f 6 60 times min b I E Ratio 1 4 1 3 1 2 1 1 5 1 1 2 1 15 c Tidal volume 0 80 1200mL 15 d Ventilating volume 18L min 1 1 4 Monitor Performanc...

Page 6: ...product should be stored in the room with temperature of 0 40 C relative humidity of no more than 80 good ventilation and without corrosion gas Note When the storage conditions are beyond the requirem...

Page 7: ...6300A Operator s Manual 3 NewTech Medical Limited 2 SYSTEM OPERATION PRINCIPLE 2 1 System Block diagram Figure 2 1 6300A Diagram...

Page 8: ...ue of the safety valve 5 1kPa 6kPa the safety valve can open automatically and the gas will be discharged from the safety valve The electromagnetic valve will be closed in expiration the pressure exer...

Page 9: ...Fig 4 Table1 Figure 2 3 Table 2 1 Display symbol s implication Function TIDAL VOL Display the tidal volume value AIRWAY PRESSURE Airway pressure display Reflect the airway pressure change when patien...

Page 10: ...airway pressure limit setting key LO PRESS Low airway pressure limit setting key POWER Power Indicator 2 2 3 Function key Figure 2 5 Table 2 3 Symbol Function A C Intermittent positive pressure ventil...

Page 11: ...Airway High Pressure Alarm Display LO PRESS Airway Low Pressure Alarm Display Apnea Apnea Alarm Tidal Volume Tidal volume regulating knob Regulate ranging 0 80ml 1200ml 2min Eliminate alarm When alarm...

Page 12: ...ilator Component Figure 3 1 1 Ventilator 2 Machine Arm 3 Power Cable For Ventilator 4 Check Valve 5 Oxygen Pipeline 6 Cart 7 Wheels 8 Bag 9 Water Trap 10 Lines For Measure Temperature 11 Pressure Samp...

Page 13: ...power supply should be used The fuse of the 6300A ventilator is at the power socket position of the rear panel The fuse with the value of 2A can be replaced after the power cable is unplugged and the...

Page 14: ...essure check Set the trigger pressure at 0 1cm H2O Pinch the simulated lung with both hands When the airway pressure is a little lower than the preset value trigger inspiration and the indicating ligh...

Page 15: ...the key flashes and the revised part flashes At this moment press or key and set the respiratory frequency HI PRESS Airway upper pressure limit setting When pressing HI PRESS key the light at the lef...

Page 16: ...gh When pressing the key the light at the left upper corner of the key flashes During A C every certain time provide double tidal volume once SIMV f 2 Synchronous intermittent mandatory ventilation Wh...

Page 17: ...igher than the airway pressure peak value Figure 5 1 Control Ventilation Waveform 5 2 A Namely assist respiration It is featured by that the patient can control the respiration frequency while the tid...

Page 18: ...ence breathe when Other parameter settings are same as that of A C b SIMV f 4 The command ventilation frequency is one fourth of A C ventilation frequency Others are same as that of f 2 Figure 5 3 SIM...

Page 19: ...tient have spontaneous respiration select SPONT should be provided for the patient through facemask before offline so as to train the spontaneous respiration ability of the patient As a routine the me...

Page 20: ...er to the above for the disinfectant method But the flow sensor should be paid special attention as it is precision and damageable part For disinfectant it is better to soak it in alcohol for an hour...

Page 21: ...irection and it should not be installed in reverse Also the sensor of the inspiration circuit and that of the expiration circuit can t be mistaken Generally the expiration channel is easy to be pollut...

Page 22: ...operation therapy or nursing of the patient Special the airway pipeline and power supply cable on the ground should not affect walking of person so as to avoid them being contacted and causing gas and...

Page 23: ...off Switch on the power source Digital number tube doesn t flash Ventilator doesn t work Fuse melts Replace During operation ventilator stops suddenly indicating light goes out with sound alarm Power...

Page 24: ...defects for a period of eighteen 18 months and accessories for a period of three 3 months The warranty period starts at the date of purchase The warranty is void if the ventilator machines have not b...

Reviews: