6300A Operator’s Manual
NewTech Medical Limited - 12 -
4.2 Operation
Table 4-2
Parameter key
Operating method
A/C
Control ventilation:
When pressing the key, the light at the left upper corner of the key
flashes. At this mode, in case of the patients’ respiratory condition,
ventilate for the patients according to the pre-adjusted ventilating
parameters.
Assist ventilation:
When pressing the key, the light at the left upper corner of the key
flashes. At this mode, inspiration depends on patients. Other parameters
ventilate for the patients according to the pre-adjusted ventilating
parameters.
A/C+ SIGH
Sigh:
When pressing the key, the light at the left upper corner of the key
flashes. During A/C, every certain time provide double tidal volume
once.
SIMV f/2
Synchronous intermittent mandatory ventilation:
When pressing the key, the light at the left upper corner of the key
flashes. Patient controls the spontaneous respiration frequency, but the
value is one-second of the preset A/C frequency.
SIMVf/4
Synchronous intermittent mandatory ventilation:
When pressing the key, the light at the left upper corner of the key
flashes. Patient controls the spontaneous respiration frequency, but the
value is one-fourth of the preset A/C frequency.
SPONT
When pressing the key, the light at the left upper corner of the key
flashes. The patient spontaneous respiration via continuous positive
pressure gas flow system.