      - 11 -                                                              NewTech Medical Limited 

 

 

4 OPERATION 

This chapter will mainly introduce the detailed operating method of 6300A ventilator. 
 

4.1 Parameters’ 

setting 

Table 4-1

 

Parameter key 

Operating method 

Tidal Volume 

Turn the knob to change the tidal volume value. Increase by clockwise 
and decrease by counter-clockwise. 

I: E 

Inspiration and respiratory ratio setting: 
When pressing I: E key, the light at the left upper corner of the key flashes 
and the revised part flashes. At this moment press 

Δ

 or 

 key, and set the 

I: E mode. 

RATE 

Respiratory frequency setting: 
When pressing RATE key, the light at the left upper corner of the key 
flashes and the revised part flashes. At this moment press 

Δ

 or 

 key, and 

set the respiratory frequency. 

HI PRESS   

Airway upper pressure limit setting: 
When pressing HI PRESS key, the light at the left upper corner of the key 
flashes and the revised part flashes. At this moment press 

Δ

 or 

 key, and 

set the airway upper pressure limit. 

LO PRESS 

Airway lower pressure limit setting: 
When pressing LO PRESS key, the light at the left upper corner of the 
key flashes and the revised part flashes.    At this moment press 

Δ

 or 

key, and set the airway lower pressure limit. 

TRIG 

Trigger pressure setting: 
When pressing TRIG key, the light at the left upper corner of the key 
flashes and the revised part flashes. At this moment press 

Δ

 or 

 key, and 

change the trigger pressure value. 

2min 

Eliminate alarm 
When alarming, press the key to eliminate alarm. 

 

 

Summary of Contents for 6300A

Page 1: ...L VENTILATOR 6300A NewTech Medical Limited 8400 Normandale Lake Boulevard Suite 920 Bloomington 55437 Minnesota U S A Tel 1 952 837 2688 Fax 1 952 400 8947 http www newtech medical com E mail info new...

Page 2: ...LATION AND CHECK 8 3 1 Ventilator Component 8 3 2 Ventilator Installation 9 3 3 Check 9 4 OPERATION 11 4 1 Parameters setting 11 4 2 Operation 12 5 OPERATION METHODS OF VARIOUS VENTILATION MODES 13 5...

Page 3: ...m the machine Always use a fresh bacterial filter and test circuit A hazard to the health of the service person may result z Hazardous voltages exist within this medical device Always observes appropr...

Page 4: ...before putting the machine into operation and strictly follow all the procedures of operation Our company guarantees the quality of the product Any time and in any case when you need our help we are...

Page 5: ...PONT 1 1 3 Ventilation Performance a Respiration frequency f 6 60 times min b I E Ratio 1 4 1 3 1 2 1 1 5 1 1 2 1 15 c Tidal volume 0 80 1200mL 15 d Ventilating volume 18L min 1 1 4 Monitor Performanc...

Page 6: ...product should be stored in the room with temperature of 0 40 C relative humidity of no more than 80 good ventilation and without corrosion gas Note When the storage conditions are beyond the requirem...

Page 7: ...6300A Operator s Manual 3 NewTech Medical Limited 2 SYSTEM OPERATION PRINCIPLE 2 1 System Block diagram Figure 2 1 6300A Diagram...

Page 8: ...ue of the safety valve 5 1kPa 6kPa the safety valve can open automatically and the gas will be discharged from the safety valve The electromagnetic valve will be closed in expiration the pressure exer...

Page 9: ...Fig 4 Table1 Figure 2 3 Table 2 1 Display symbol s implication Function TIDAL VOL Display the tidal volume value AIRWAY PRESSURE Airway pressure display Reflect the airway pressure change when patien...

Page 10: ...airway pressure limit setting key LO PRESS Low airway pressure limit setting key POWER Power Indicator 2 2 3 Function key Figure 2 5 Table 2 3 Symbol Function A C Intermittent positive pressure ventil...

Page 11: ...Airway High Pressure Alarm Display LO PRESS Airway Low Pressure Alarm Display Apnea Apnea Alarm Tidal Volume Tidal volume regulating knob Regulate ranging 0 80ml 1200ml 2min Eliminate alarm When alarm...

Page 12: ...ilator Component Figure 3 1 1 Ventilator 2 Machine Arm 3 Power Cable For Ventilator 4 Check Valve 5 Oxygen Pipeline 6 Cart 7 Wheels 8 Bag 9 Water Trap 10 Lines For Measure Temperature 11 Pressure Samp...

Page 13: ...power supply should be used The fuse of the 6300A ventilator is at the power socket position of the rear panel The fuse with the value of 2A can be replaced after the power cable is unplugged and the...

Page 14: ...essure check Set the trigger pressure at 0 1cm H2O Pinch the simulated lung with both hands When the airway pressure is a little lower than the preset value trigger inspiration and the indicating ligh...

Page 15: ...the key flashes and the revised part flashes At this moment press or key and set the respiratory frequency HI PRESS Airway upper pressure limit setting When pressing HI PRESS key the light at the lef...

Page 16: ...gh When pressing the key the light at the left upper corner of the key flashes During A C every certain time provide double tidal volume once SIMV f 2 Synchronous intermittent mandatory ventilation Wh...

Page 17: ...igher than the airway pressure peak value Figure 5 1 Control Ventilation Waveform 5 2 A Namely assist respiration It is featured by that the patient can control the respiration frequency while the tid...

Page 18: ...ence breathe when Other parameter settings are same as that of A C b SIMV f 4 The command ventilation frequency is one fourth of A C ventilation frequency Others are same as that of f 2 Figure 5 3 SIM...

Page 19: ...tient have spontaneous respiration select SPONT should be provided for the patient through facemask before offline so as to train the spontaneous respiration ability of the patient As a routine the me...

Page 20: ...er to the above for the disinfectant method But the flow sensor should be paid special attention as it is precision and damageable part For disinfectant it is better to soak it in alcohol for an hour...

Page 21: ...irection and it should not be installed in reverse Also the sensor of the inspiration circuit and that of the expiration circuit can t be mistaken Generally the expiration channel is easy to be pollut...

Page 22: ...operation therapy or nursing of the patient Special the airway pipeline and power supply cable on the ground should not affect walking of person so as to avoid them being contacted and causing gas and...

Page 23: ...off Switch on the power source Digital number tube doesn t flash Ventilator doesn t work Fuse melts Replace During operation ventilator stops suddenly indicating light goes out with sound alarm Power...

Page 24: ...defects for a period of eighteen 18 months and accessories for a period of three 3 months The warranty period starts at the date of purchase The warranty is void if the ventilator machines have not b...

Reviews: