6 CLEANING 

AND 

MAINTENANCE 

 

6.1 General 

principle 

Generally, the gas in the inspiration circuit (the part before the humidifier) is dry and clean, and 
flows to the patient with a certain velocity, so it is unsuitable for the bacteria to grow. While the gas 
in the expiration circuit is from patient, it is humid, and some patients have infectious disease, 
which is easy for the bacteria to increase, so the cleaning and disinfectant operations are specified 
in principle as follows. 

 

6.1.1 Non-infective 

patient 

Just conduct disinfectant for the connectors such as patient circuit (silicon rubber threaded pipe), 
water trap and Y joint out of the airway for each patient. The disinfectant method can be fumigating, 
infection liquid soaking or high-temperature sterilization, decided in accordance with the hospital's 
conditions. 

 

6.1.2 Infective 

patient 

Besides conducting disinfectant for the patient circuit, the gas delivery element (in the airway tank) 
of the expiration circuit should also be disinfected so as to avoid cross infestation between patients 
and prevent the ward from being polluted so that the health of staffs is affected. Refer to the above 
for the disinfectant method. But the flow sensor should be paid special attention, as it is precision 
and damageable part. For disinfectant, it is better to soak it in alcohol for an hour, and then dry it. 
While soaking, the plug part of the sensor should be upward and exposed in the air, it will be OK if 
the alcohol height can drown the ventilation hole. When positioning the disinfected sensor, pay 
attention to positioning the side marked with arrowhead upward (the arrow head means the flow 
direction). 
 

6.1.3 

Surface cleaning and sterilization 

Clean the machine's panel and all surfaces with soft cloth soaked with the soluble sterilizing agent. 
The formation and preparation of the agent must be done in accordance with the direction given by 
the manufacturer. One must prevent the sterilizing agent drops from entering the anesthetic 
machine and the misusage of organic agent for cleaning the machine. 
 

6.1.4 Expiratory 

valves 

cleaning 

Dismount the cover of the expiratory valves by rotating it counter clockwise, then clean all parts of 
them with the gauze socked with water soluble sterilizing agent, after all parts cleaned and dried 
recover it in original integration. Then one must check the leakage and the movement of the 
expiratory valves in accordance with the required regulation and checking procedure. Please 
handle all parts with care preventing any damage. 

Summary of Contents for 6300A

Page 1: ...L VENTILATOR 6300A NewTech Medical Limited 8400 Normandale Lake Boulevard Suite 920 Bloomington 55437 Minnesota U S A Tel 1 952 837 2688 Fax 1 952 400 8947 http www newtech medical com E mail info new...

Page 2: ...LATION AND CHECK 8 3 1 Ventilator Component 8 3 2 Ventilator Installation 9 3 3 Check 9 4 OPERATION 11 4 1 Parameters setting 11 4 2 Operation 12 5 OPERATION METHODS OF VARIOUS VENTILATION MODES 13 5...

Page 3: ...m the machine Always use a fresh bacterial filter and test circuit A hazard to the health of the service person may result z Hazardous voltages exist within this medical device Always observes appropr...

Page 4: ...before putting the machine into operation and strictly follow all the procedures of operation Our company guarantees the quality of the product Any time and in any case when you need our help we are...

Page 5: ...PONT 1 1 3 Ventilation Performance a Respiration frequency f 6 60 times min b I E Ratio 1 4 1 3 1 2 1 1 5 1 1 2 1 15 c Tidal volume 0 80 1200mL 15 d Ventilating volume 18L min 1 1 4 Monitor Performanc...

Page 6: ...product should be stored in the room with temperature of 0 40 C relative humidity of no more than 80 good ventilation and without corrosion gas Note When the storage conditions are beyond the requirem...

Page 7: ...6300A Operator s Manual 3 NewTech Medical Limited 2 SYSTEM OPERATION PRINCIPLE 2 1 System Block diagram Figure 2 1 6300A Diagram...

Page 8: ...ue of the safety valve 5 1kPa 6kPa the safety valve can open automatically and the gas will be discharged from the safety valve The electromagnetic valve will be closed in expiration the pressure exer...

Page 9: ...Fig 4 Table1 Figure 2 3 Table 2 1 Display symbol s implication Function TIDAL VOL Display the tidal volume value AIRWAY PRESSURE Airway pressure display Reflect the airway pressure change when patien...

Page 10: ...airway pressure limit setting key LO PRESS Low airway pressure limit setting key POWER Power Indicator 2 2 3 Function key Figure 2 5 Table 2 3 Symbol Function A C Intermittent positive pressure ventil...

Page 11: ...Airway High Pressure Alarm Display LO PRESS Airway Low Pressure Alarm Display Apnea Apnea Alarm Tidal Volume Tidal volume regulating knob Regulate ranging 0 80ml 1200ml 2min Eliminate alarm When alarm...

Page 12: ...ilator Component Figure 3 1 1 Ventilator 2 Machine Arm 3 Power Cable For Ventilator 4 Check Valve 5 Oxygen Pipeline 6 Cart 7 Wheels 8 Bag 9 Water Trap 10 Lines For Measure Temperature 11 Pressure Samp...

Page 13: ...power supply should be used The fuse of the 6300A ventilator is at the power socket position of the rear panel The fuse with the value of 2A can be replaced after the power cable is unplugged and the...

Page 14: ...essure check Set the trigger pressure at 0 1cm H2O Pinch the simulated lung with both hands When the airway pressure is a little lower than the preset value trigger inspiration and the indicating ligh...

Page 15: ...the key flashes and the revised part flashes At this moment press or key and set the respiratory frequency HI PRESS Airway upper pressure limit setting When pressing HI PRESS key the light at the lef...

Page 16: ...gh When pressing the key the light at the left upper corner of the key flashes During A C every certain time provide double tidal volume once SIMV f 2 Synchronous intermittent mandatory ventilation Wh...

Page 17: ...igher than the airway pressure peak value Figure 5 1 Control Ventilation Waveform 5 2 A Namely assist respiration It is featured by that the patient can control the respiration frequency while the tid...

Page 18: ...ence breathe when Other parameter settings are same as that of A C b SIMV f 4 The command ventilation frequency is one fourth of A C ventilation frequency Others are same as that of f 2 Figure 5 3 SIM...

Page 19: ...tient have spontaneous respiration select SPONT should be provided for the patient through facemask before offline so as to train the spontaneous respiration ability of the patient As a routine the me...

Page 20: ...er to the above for the disinfectant method But the flow sensor should be paid special attention as it is precision and damageable part For disinfectant it is better to soak it in alcohol for an hour...

Page 21: ...irection and it should not be installed in reverse Also the sensor of the inspiration circuit and that of the expiration circuit can t be mistaken Generally the expiration channel is easy to be pollut...

Page 22: ...operation therapy or nursing of the patient Special the airway pipeline and power supply cable on the ground should not affect walking of person so as to avoid them being contacted and causing gas and...

Page 23: ...off Switch on the power source Digital number tube doesn t flash Ventilator doesn t work Fuse melts Replace During operation ventilator stops suddenly indicating light goes out with sound alarm Power...

Page 24: ...defects for a period of eighteen 18 months and accessories for a period of three 3 months The warranty period starts at the date of purchase The warranty is void if the ventilator machines have not b...

Reviews: