December 2019
105
LDC-3706 Series Laser Controller
9. The instrument will ramp to a new setpoint. Repeat step 7 for the newly calculated
photodiode current at this setpoint.
10. Wait for the 7-segment display to exit the calibration mode. Once the self-calibration is
completed, the calibration constants will be stored to the non-volatile memory and the
display will return to its previous state.