December 2019
108
LDC-3706 Series Laser Controller
Figure A.1 – Non Linearity Graph
If a maximum absolute error of 0.8°C is tolerable (over the entire temperature range), the one
point calibration of C
1
should be used. If C
1
is calibrated at 25°C, and the intended operating
range is 0 to 50°C, a maximum error of about ±0.2°C may be expected over that operating
range. If a greater accuracy is desired, the two point method of determining C
1
and C
2
should be
used. Note however, the absolute error curve is non-linear; therefore, the constant C
2
will vary
over different temperature ranges.
LM335 Sensor
The LM335 is a linear thermal sensor which acts as a constant voltage regulator. It produces a
voltage, V, which is directly proportional to absolute temperature, over its useful range (-40°C to
+100°C). This nominal value can be expressed as:
= 10
Where V is the nominal voltage produced by the LM335 and K is the temperature in Kelvin.
The LDC-3706 instruments use V to determine the nominal temperature, T
n
, by the formula:
=
10
−
273.15
Where T
n
is measured in
°
C.
The temperature, T
d
, which is displayed by the LDC-3706, is first calibrated as follows:
=
+
10
∗
Where C
1
and C
2
are the constants stored by the user in the LDC-3706 for the LM335.
Summary of Contents for LDC-3706 Series
Page 2: ......
Page 12: ...December 2019 x LDC 3706 Series Laser Controller...
Page 18: ...December 2019 6 LDC 3706 Series Laser Controller LDC 3706 Specifications...
Page 19: ...December 2019 7 LDC 3706 Series Laser Controller...
Page 20: ...December 2019 8 LDC 3706 Series Laser Controller...
Page 21: ...December 2019 9 LDC 3706 Series Laser Controller...
Page 22: ...December 2019 10 LDC 3706 Series Laser Controller...
Page 52: ...December 2019 40 LDC 3706 Series Laser Controller...
Page 64: ...December 2019 52 LDC 3706 Series Laser Controller...
Page 101: ...December 2019 89 LDC 3706 Series Laser Controller...
Page 118: ...December 2019 106 LDC 3706 Series Laser Controller...