6
Other Data
Maximum current automatic switches for the feeding
net
Magnetic thermal switch from 10 A (CEI
23-3)
High voltage measure method:
Not invasive method
Current measure method in the radiogenic pipe:
See page 54
Charge application time determination method:
Not invasive method
Interposition aluminium filter between rays window and
cone collimator:
Al 1 mm (AIP99,9 UNI3567)
INSTALLATION AND USE
Use Conditions
The equipment is designed for continuous operation with intermittent charge.
The operating times are with intermittent charge with a duty cycle of 1.30 (for each
working period will correspond to 30 pause)
Classification according to Directive 2007/47 / EC
(Legislative Decree no. 37/2010)
Class IIb
Protection against electrical dangers:
Class I
Protection degree against the direct and indirect
contacts:
Equipment with parts applied type B
Protection degree against water penetration:
Common equipment IPX0
Use safety degree in presence of inflammable
anaesthetic mixture:
Equipment not suitable to an use in
presence of an inflammable anaesthetic
mixture with air or with oxygen or with
nitrogen protoxide
Use conditions:
Equipment for continuous operation with
intermittent charge
Installation:
Permanent e mobile