Manual
Doc. HBE220-1
Installation and User Manual BEST-X-DC WITH
TIM-X
Mod. MQI006-0
NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected]
www.newliferadiology.it
71
Possible drawbacks found in the radiographs results
Clear image
Possible causes:
Exhausted development liquid
Excessive dilution of the developing liquid
Too short exposure time to X-rays
Insufficient development time
Exposure liquid temperature below the recommended range
Dark image
Possible causes:
Wrong dilution of developing liquid
Too long exposure time to X-rays
Excessive development time
Exposure liquid temperature higher than the recommended
Not detailed image
Possible causes:
Animal movement
Monoblock movement
Radiography partially exposed
Possible causes:
Error in the centering between rays’s beams and film
Liquid development too low with a consequent film partial development
Contact between two or more films during development
Veiled Image
Possible causes:
Films that exceeded the expiration date
Film accidental exposure to rays
Film accidental exposure to heat sources
Film accidental exposure to daylight or safety lamp darkroom no longer suitable
Black line occurrence on X-rays
Possible causes: