Manual
Doc. HBE220-1
Installation and User Manual BEST-X-DC WITH
TIM-X
Mod. MQI006-0
NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected]
www.newliferadiology.it
10
Information for installation
PLASTIC MATERIAL PLUGS NOT ALLOWED
Utensils and tools required for installation (not supplied)
1 multi meter
1 meter
1 fixed key from 13 mm
1 nandle spanner from 5,5 mm
1 set of Allen spanner
1 spirit level
1 plastic mallet
1 percussion drill with points from Ø 3 to Ø 13 mm
1 thin screwdriver for electrical connection
1 medium screwdriver
1 net feeding cable with three wires (2 cond 1 ground) of section 1,5 mmq for
a reduced length of not over 40 m (for superior lengths use the section cable
corresponding to that indicated from the Country actual standards).
Electrical indications
All works for the feeding electric plant must be performed in conformance with the actual
reference standards for plants and locals for medical use. (CEI 64-8-710).
A feeding of 230 V~ to 50 Hz. is required. For the phase conductors, neutral and ground, the
minimum section must be of 1,5 mmq of copper. It recalls the necessity to perform the
ground connection as required by current legislation.
The plant must be performed and tested by qualified personnel.
The warranty excludes damages caused by erroneous connection.
Wiring and connection conductors sections
A three conductors cable (3x1,5 mmq) is expected between the net switch and the
equipment. For the assembled versions and in all cases of separate installation from the
control unit, a second cable, always 3 x 1.5 sq mm, shall be provided between the control
unit and the support. Please note that the radiological groups are supplied without a plug