Installation and User Manual BEST-X-DC WITH
TIM-X
Mod. MQI006-0
NEW LIFE RADIOLOGY S.R.L. Corso Allamano 13/15 interno G 10095 Grugliasco (TO)
Italy phone and fax + 39.011. 781994 [email protected]
www.newliferadiology.it
3
Palmtop
Palmtop
of
the
BEST-X-DC
allows
managing the exposure times and
adjustment of kV(60/70) and mA (4/7) and
at the same time it guarantees maximum
safety in the use of X-ray tubes for
electromedical purposes for intraoral
diagnostic;
The management of the palmtop occurs
through the use of the RF radio control;
The control panel of the palmtop is
provided with touchscreen display so as to
facilitate its use;
The functioning of the palmtop is
managed by RF radio control to vigilant
device better known as dead man’s
device, in such a way as to guarantee
maximum safety for both, the operator
and the patient;
The palmtop does not permit use in the X-
ray scopy;
The time scale set by the values of the
factory complies with scale R10 relative to
the standard EN60601-2-7 (IEC 60601-2-7);
The exposure times shown on the display of
the timer are expressed in ms.
The maximum time that can be set is 1 s
(the display will read 1000) while the
minimum time is 20 ms. (the display will
read 0020);