X-Ray Pre-Start
In this screen the hand-held seeks the receiver and, once found it, programs it making
“Configuration done” appear.
At this point it is possible to do the x-ray pressing the X-Ray button and keeping it pressed.
If the button is pressed prior to the configuration the x-ray does not occur and the button
releasing is requested, if instead, the receiver is not found, after a certain period of time it is
necessary to exit such screen.
To exit, in any case, it is sufficient to press the Cancel button.
To carry out an X-ray, you need to keep the X-Ray button pressed until the cooling process
screen appears.