10
New Brunswick Scientific
User’s Guide
2
2
S
S
P
P
E
E
C
C
I
I
F
F
I
I
C
C
A
A
T
T
I
I
O
O
N
N
S
S
Innova 4200 Incubator Shaker
construction
Heavy gauge stainless steel
door
26 3/8
″
H x 22
″
W (670 mm x 559 mm)
Cabinet double
paned
glass window
19¼
″
H x 11 3/8
″
W (489 mm x 289 mm)
fluorescent
light
13 watt–provides illumination inside chamber
shelves
2 Stainless Steel 20
″
W x 20
″
L (508 mm x 508 mm)–adjustable–
(½
″
(12.7 mm) centers)
internal
AC
receptacle
200W AC–waterproof w/ mating connector–can be used w/
magnetic stirrer, spinner flask, rocker, roller & shakers
platform
18
″
W x 18
″
L (457 mm x 457 mm)
Dimensions
overall
†
24.4
″
W x 28.7
″
L x 36.4
″
H (620 mm x 729 mm x 925 mm)
chamber
20½
″
W x 20½
″
L x 24.8
″
H (520 mm x 520 mm x 630 mm)
Electrical
1200 Watts–120/100/200/240V–50/60Hz
Indicator
3 digit LED
Recorder Output
Two 0-5 Volt–tracks temperature & speed w/ optional recorder
alarm
Audible & visible indicates when speed deviates
±
5 rpm–audible
alarm can be deactivated
control
±
1 rpm–acceleration control minimizes splashing caused by
sudden starts and stops
display
1 rpm increments
drive
Triple-eccentric counterbalanced drive w/ 9 sealed for life, ball
bearings
Shaking motion
¾
″
(19 mm) circular–standard
1
″
(25.4) optional
motor
1/15 HP, 3-phase solid-state, brushless ball bearing DC motor
range
25–400 rpm, 25-300 rpm for stacked unit
timer
Programmable shaking continuous
→
99.9 hours–setable in 0.1
hr. increments–alarm sounds at end of shaking period
automatic
restart
Setpoints & operating status retained by non-volatile memory–
auto restart after power is restored
door
interlock
Shaker shuts off when door is open
Safety unbalanced/
overload sensor
Limits speed automatically when unbalanced or overload
condition is detected
chamber
temperature
Heater shuts off if temperature exceeds operating range
†
51.2 (130 cm) front to back w/ door open
Summary of Contents for innOva 4200
Page 2: ...2 New Brunswick Scientific User s Guide...
Page 57: ...57 Innova 4200 M1233 0053 User s Guide Figure 21 Electrical Schematic Quadrant A...
Page 58: ...58 New Brunswick Scientific User s Guide Figure 22 Electrical Schematic Quadrant B...
Page 59: ...59 Innova 4200 M1233 0053 User s Guide Figure 23 Electrical Schematic Quadrant C...
Page 60: ...60 New Brunswick Scientific User s Guide Figure 24 Electrical Schematic Quadrant D...
Page 66: ...66 New Brunswick Scientific User s Guide Figure 30 100V Power Schematic Overview...
Page 67: ...67 Innova 4200 M1233 0053 User s Guide Figure 31 100V Power Schematic Quadrant A...
Page 68: ...68 New Brunswick Scientific User s Guide Figure 32 100V Power Schematic Quadrant B...
Page 69: ...69 Innova 4200 M1233 0053 User s Guide Figure 33 100V Power Schematic Quadrant C...
Page 70: ...70 New Brunswick Scientific User s Guide Figure 34 100V Power Schematic Quadrant D...