45
Innova 4200
•
M1233-0053
User’s Guide
9.7
Temperature Sensor Adjustment
The temperature sensor is located in the bottom right of the center panel of the back wall.
It is held in place by a bracket and can be adjusted vertically. The sensor should be
adjusted so that there is a ½ inch gap between the tip of the sensor and the top of the slot
that it rests in front of.
Figure 17: Temperature Sensor Adjustment
To adjust the temperature sensor:
1. Set
the
ON /OFF POWER SWITCH
to
OFF
.
2. Disconnect
the
POWER CORD
from the electrical outlet.
3. Remove all the shelving from the inside cabinet.
4.
If the sensor and bracket have been removed,
set the sensor, sensor bracket and
mounting screws in place, but do not secure.
5.
If the sensor is still in place,
loosen the sensor bracket screws to allow the sensor to
move. Adjust the sensor so that the gap between the tip of the sensor and the top of
the slot is ½ inch (
see Figure 17 above
). Secure the sensor bracket screws in place.
1/2”
Summary of Contents for innOva 4200
Page 2: ...2 New Brunswick Scientific User s Guide...
Page 57: ...57 Innova 4200 M1233 0053 User s Guide Figure 21 Electrical Schematic Quadrant A...
Page 58: ...58 New Brunswick Scientific User s Guide Figure 22 Electrical Schematic Quadrant B...
Page 59: ...59 Innova 4200 M1233 0053 User s Guide Figure 23 Electrical Schematic Quadrant C...
Page 60: ...60 New Brunswick Scientific User s Guide Figure 24 Electrical Schematic Quadrant D...
Page 66: ...66 New Brunswick Scientific User s Guide Figure 30 100V Power Schematic Overview...
Page 67: ...67 Innova 4200 M1233 0053 User s Guide Figure 31 100V Power Schematic Quadrant A...
Page 68: ...68 New Brunswick Scientific User s Guide Figure 32 100V Power Schematic Quadrant B...
Page 69: ...69 Innova 4200 M1233 0053 User s Guide Figure 33 100V Power Schematic Quadrant C...
Page 70: ...70 New Brunswick Scientific User s Guide Figure 34 100V Power Schematic Quadrant D...