29
Innova 3100
M1231-0050
User’s Guide
10.
Mount the new motor (part #M1195-4000) to the motor plate with the three flat
head screws (from step 8).
11.
Replace the pulley. Screw the set screw to the shaft flat, but do not tighten.
12.
Position this assembly back onto the unit. Replace the two hex head bolts with
their associated hardware, but do not tighten.
13.
Replace the belt. Adjust the motor pulley height so that the belt is level as related
to the drive pulley, then tighten the set screw.
14.
Adjust the belt tension as shown in
Figure 9
, then tighten the two hex head bolts.
15.
Rotate the large pulley by hand and see that the belt tracks smoothly.
16.
Replace the motor connector. Be sure the motor connector is positioned (with the
pin visible) as shown in
Figure 10
.
17.
Reinstall the ground lug, lock washer and #8-32 nut onto the stud on the motor
plate.
18.
Reinstall the bottom cover with its Phillips head screws.
19.
Reconnect the power cord to the unit and to the power source.
20.
Set the
ON/OFF
switch to the ON position.
The unit is ready for operation.
5.7
Acknowledging the MAINT Indicator
After the shaker has been operating for 10,000 hours accumulated running time
(initially, or since the previous maintenance check), the MAINT indicator light on the
control panel will light. Only an NBS service engineer can turn it off.
The MAINT light indicates that it is time for a routine maintenance check. A regular
schedule of routine maintenance is an excellent way to keep your valuable equipment
performing optimally for years of reliable service.
5.8 Recorder
Adaptation
To record speed and temperature a recorder (not supplied) can be used. It should have
the following capabilities:
•
To record both speed and temperature, two channels are required. Each
channel should have signal conditioning which accepts 0-5 volt DC input. The
pin-out diagram and scale below identify the application.
•
A mating connector is required on the recorder cable (not supplied). This is a
9-pin male D subminiature connector, AMP Amplimite HDP-20 series or
equivalent.
Summary of Contents for innova 3100
Page 2: ...ii New Brunswick Scientific User s Guide...
Page 4: ...iv New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 6: ...vi New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 8: ...viii New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 10: ...x New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 14: ...2 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 16: ...4 New Brunswick Scientific User s Guide Figure 1 Front View...
Page 28: ...16 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 36: ...24 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 48: ...36 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 54: ...42 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 55: ...43 Innova 3100 M1231 0050 User s Guide 8 8 D DR RA AW WI IN NG GS S 8 1 Control Schematic...