20
New Brunswick Scientific
User’s Guide
2.
Increase or decrease the setpoint by pressing the
∆
or
∇
key.
The temperature alarms, both audible and flashing light, are activated if the
temperature is more than 1.0
°
C higher or lower than the temperature setpoint.
The alarm will automatically deactivate as the unit achieves the set
temperature.
4.6.2 Deactivating
Temperature Control
If desired, the temperature control system may be shut off during set-up for
special investigations.
To deactivate the temperature control system:
1.
Press and hold the
∇
key until the setpoint is at 4.0
°
C.
2.
While holding the
∇
key, simultaneously press the START/STOP key.
The temperature setpoint display shows “Off” and the heater (and the
cooling coil, if present) will be deactivated.
To reactivate the temperature control:
•
Press the
∆
key until the desired temperature setpoint is displayed.
#
NOTE:
The shaker may be started or stopped by pressing the
START/STOP
key.
When starting, the unit will automatically return to the last function and
setting. The alarms will be activated until the speed is within 5 RPM or
the temperature is within 1.0
°
C of the setpoint. The alarm will not sound
when the shaker is accelerating or the temperature is changing to satisfy
the setpoint immediately following turning on the power.
4.7
Temperature Offset Calibration
The temperature probe and the temperature controller are calibrated together at the
factory. The temperature probe measures the temperature of the air at the probe’s
location, near the heat exchanger return vent. The controller uses the probe input to
adjust air temperature, up or down, to match the temperature setpoint.
Summary of Contents for innova 3100
Page 2: ...ii New Brunswick Scientific User s Guide...
Page 4: ...iv New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 6: ...vi New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 8: ...viii New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 10: ...x New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 14: ...2 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 16: ...4 New Brunswick Scientific User s Guide Figure 1 Front View...
Page 28: ...16 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 36: ...24 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 48: ...36 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 54: ...42 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 55: ...43 Innova 3100 M1231 0050 User s Guide 8 8 D DR RA AW WI IN NG GS S 8 1 Control Schematic...