4.4.2
Cancelling the Timer
To cancel the timer WITHOUT stopping the shaker:
1.
Press the
SELECT
key to light HRS.
2.
Set the time by pressing the
∆
or
∇
key until the desired setpoint is
displayed (0.1 - 99.9 hours).
3.
While the SET LED is lit, press the
START/STOP
key. The TIME
indicator will go out and the display will read “Off”.
4.5 Alarm
Functions
Innova shakers have an audible alarm which is activated at predetermined times. It
may be deactivated in the following way:
1.
Press
SELECT
to light HRS.
2.
Simultaneously
press the
∆
and
∇
keys. The SET and MAINT indicators will
flash.
3.
While the SET and MAINT indicators are flashing press the START/STOP key.
The MUTE indicator will light to advise that the audible alarm is deactivated.
To reactivate the alarm, repeat steps 1-3. The MUTE indicator will be extinguished
when the alarm has been reactivated.
4.6 Temperature
This control consists of an internal electrical interface, an RTD temperature probe, and
an analog output for chart recorder or computer. The temperature probe is in the
water bath. The temperature can be set from 5.0
°
C above ambient to 99.9
°
C.
4.6.1
Setting the Temperature
To set the temperature setpoint:
1.
Press the
SELECT
key until the function indicator lights on the
°
C mode.
#
NOTE:
For temperatures above 50
°
C, the stainless steel gable cover (
see
Chapter 7 for the part number
) is required to minimize heat loss.
Summary of Contents for innova 3100
Page 2: ...ii New Brunswick Scientific User s Guide...
Page 4: ...iv New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 6: ...vi New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 8: ...viii New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 10: ...x New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 14: ...2 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 16: ...4 New Brunswick Scientific User s Guide Figure 1 Front View...
Page 28: ...16 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 36: ...24 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 48: ...36 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 54: ...42 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 55: ...43 Innova 3100 M1231 0050 User s Guide 8 8 D DR RA AW WI IN NG GS S 8 1 Control Schematic...