User’s Guide
5.6
Replacing the Motor Assembly
To replace the motor assembly:
1.
Drain the water out of the bath.
2.
Turn off the power and disconnect the power cord.
3.
Lay the unit on its left side, using a soft pad to protect the finish.
4.
Remove the bottom cover using a Phillips (+) screwdriver. Retain the screws for
reuse.
5.
Referring to
Figure 10 below
, remove the connector from the motor by lifting
straight up.
Figure 10: Motor Assembly Replacement
6.
Remove the #8-32 nut, lock washer and ground lug from the motor plate.
7.
Remove the two hex head bolts and washers, then lift out the motor and plate
assembly.
8.
Separate the motor from the plate by removing the three flat head screws. Save
the hardware.
9.
Loosen the pulley set screw and remove the pulley from the shaft. Save the
pulley.
Summary of Contents for innova 3100
Page 2: ...ii New Brunswick Scientific User s Guide...
Page 4: ...iv New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 6: ...vi New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 8: ...viii New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 10: ...x New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 14: ...2 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 16: ...4 New Brunswick Scientific User s Guide Figure 1 Front View...
Page 28: ...16 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 36: ...24 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 48: ...36 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 54: ...42 New Brunswick Scientific User s Guide THIS PAGE INTENTIONALLY BLANK...
Page 55: ...43 Innova 3100 M1231 0050 User s Guide 8 8 D DR RA AW WI IN NG GS S 8 1 Control Schematic...