36
8.2 Setting
Frequency
The C-24KC is set to the appropriate frequency prior to shipment. The
FREQUENCY
SWITCH
(see drawing below)
is on the back panel of the shaker. If you relocate the
shaker and need to reset to the frequency, perform the following steps.
CAUTION!
If the C-24KC is being operated in Europe, CE Labeling requires
that the following procedure be performed only by a qualified
Electrical or Service Engineer.
1. Verify that the shaker is disconnected from the power source.
2. Slide the small red switch up for 50 Hz.
3. Slide the small red switch down for 60Hz.
Figure 6: Frequency Selector Switch
8.3 Fuse
Replacement
There are three fuses for the C-24KC, housed behind the front panel. To replace a fuse,
the qualified Electrical or Service Engineer will:
1. Set
the
ON/OFF POWER
Switch to
OFF
.
2. Disconnect
the
POWER
CORD
from the electrical outlet.
3. Open the door of the shaker to gain access to the front panel. The door
must
be open
all the way.
4. Use a Phillips head (+) screwdriver to remove the three screws that secure the hinged
front panel to the housing. Gently lay down the front panel, setting aside the screws
for reuse.