Although the C-24KC shaker requires no routine mechanical
maintenance, the user should provide some routine preventive
maintenance. After the shaker has operated for more than 10,000 hours,
the
MAINT INDICATOR
on the control panel will illuminate. This means
that the shaker requires a routine maintenance check.
Preventive maintenance keeps your equipment in proper working condition. When periodically
performed, maintenance results in longer life for your equipment and reduces time lost due to
equipment failure. We suggest that you perform the maintenance procedures outlined in the
following pages.
WARNING!
Always turn off the Shaker and disconnect the power cord from
the power supply before performing any maintenance on the
unit.
7.1 Acknowledging the MAINT Light
When the
MAINT STATUS INDICATOR
light comes on, the following steps will turn the
light off:
1. Press
the
SELECT KEY
until the
HRS FUNCTION INDICATOR
illuminates.
2. Simultaneously press the
V
KEY
and the
W
KEYS
and hold for a few seconds.
The
MAINT STATUS INDICATOR
flashes.
3. Press
the
W
KEY
and the
MAINT
light will turn off.
7.2 Cleaning
Procedures
WARNING!
Always disconnect the power cord before performing any
maintenance on the Shaker.
Do not immerse the Shaker in water or any liquids.
Do not use a cleaning fluid that is corrosive or flammable on or
around the Shaker.
.