31
6.9 Maint
Indicator
After 10,000 hours of operation, the
MAINT
INDICATOR
will illuminate. Preventive
maintenance is recommended at this point. To deactivate the
MAINT
INDICATOR
, perform
the following:
1. Press
SELECT
until the
HRS
INDICATOR
is illuminated.
2. Simultaneously press the
V
and
W
KEYS
.
3. Press the
W
KEY
.
6.10 Power
Failure
In the event of a power failure, the C-24KC Benchtop Incubator Shaker is equipped with
an automatic restart function.
If the shaker was in operation prior to the power interruption, the shaker will begin to
operate at its last entered setpoint. The
LED DISPLAY
will flash, indicating that a power
failure has occurred. Press any key to cease the flashing in the display.
New Brunswick Scientific Co., Inc.
User’s Guide