30
NOTE:
The
°C
light will pulse rapidly for a short duration to indicate it is not
operating in the factory default mode. It will pulse for a longer duration and
less rapidly (with a frequency of approximately one second) to indicate
temperature is more than one degree above or below setpoint.
To return to the factory calibration:
1. Press
the
SELECT KEY
until the function
°
C INDICATOR
illuminates.
2. Simultaneously press the
V
and
W
KEYS
. The
SET
and
MAINT
INDICATORS
will
light.
3. While
the
SET
and
MAINT
INDICATORS
are illuminated, press the
START/STOP KEY
.
6.7 Temperature
Setpoint
Press the
SELECT KEY
until the function
°
C INDICATOR
illuminates. The temperature can
be set from 15
°
C below the current ambient temperature up to 60.0
°
C. Increasing or
decreasing the setpoint is accomplished with the
V
or
W
KEYS
.
During operation, if the temperature of the chamber is more than 1.0
°
C higher or lower
than the temperature setpoint, the audible and visual alarms are triggered. The visual
alarm consists of a flashing
°
C INDICATOR
. The alarms will automatically deactivate as
the unit achieves the set temperature.
6.8
Total Running Time
The control modules of the C-24KC shaker tracks the time the shaker has been “
ON
”. To
display the accumulated running time:
1. Press
SELECT
until the
HRS
INDICATOR
is illuminated.
2. Simultaneously press the
V
and
W
KEYS
.
The
SET
and
MAINT
INDICATORS
will flash and the accumulated running time will be
displayed in hundreds of hours (i.e., “02” equals 200 hours; “102” equals 10,200 hours).
This display will continue for 10 seconds and then default to the previous mode readout.
C-24KC Benchtop Incubator Shaker
•
M1281-0050
User’s Guide