![Neural Analytics Lucid M1 User Manual Download Page 69](http://html1.mh-extra.com/html/neural-analytics/lucid-m1/lucid-m1_user-manual_1676265069.webp)
Doc #: 60-00118 Rev. A
69
Specifications
Technical Information
Dimensions:
13.6 x 12.7 x 3.7 (in inches)
Weight:
12.7 lbs
Power Supply:
100
– 240 VAC, 50 to 60 Hz, 2A
IEC 60601-1
DC Input for system:
19.2V 6.3A, 120VA
Transducer Interface:
Two (2) Neural Analytics 2Mhz Transducer connections
Peripheral Interface:
Four (4) USB 3.0 connections
Headphone 3.5mm stereo connection
One (1) HDMI port (not available on older versions)
One (1) Neural Analytics Remote Control connection
Operating Software Version:
1.3.X
(X represents a number used for minor releases with no
feature changes)
Environmental Conditions:
Operating Temperature:
+15C to +60C
Storage Temperature:
0 to +70C
Ambient Pressure:
10 PSI to 15 PSI (170m to
3133m)
Relative Humidity:
5 to 95%
(non-condensing)
Expected Service Life:
5 years
Operating Mode:
Continuous operation
Doppler Frequency:
2 MHz
Modes:
2 MHz pulsed wave
Classification:
Class 2a Equipment
Pollution degree 2
Type B patient-applied parts (transducer)
IPX0 ordinary protection against ingress of water, console
IPX1 protected against vertically falling drops, transducers
Transducer Patient Contact Material:
Luran HD-20 (BASF)