LKG 610
Netech Corporation
110 Toledo St, Farmingdale, NY 11735,
Specifications
Display
: 3-1/2 Digit LCD Display
LEAKAGE CURRENT:
0-1999 Micro Amps. All Current measurements are
made through the AAMI/ANSI ES1-1985 or IEC 601-1 test load. The meter
readings correspond to the true RMS value of the current.
Accuracy:
DC to 1 KHz: 1% FS + 1 LSD
1 Khz to 100 KHz:
±
2.5% FS + 1 LSD
100 KHz to 1 MHz:
±
4% FS + 1 LSD
RESISTANCE:
0 TO 1999 Milli Ohms.
Accuracy:
+ 1 % of r 1 LSD
DEVICE CURRENT:
0 to 19.99 Amps
Accuracy:
+ 2% Full Scale + 1 LSD
Line Voltage
: 1 to 300 Volts
Accuracy:
+ 2% + 1 Volt
TEST RECEPTACLE:
Hospital Grade 110 VAC -15 Amp or 220 VAC 10 Amp.
The rocker polarity switch selects NORMAL, OFF and REVERSE polarity to the
test receptacle, and a momentary Neutral switch will open the neutral line to the test
receptacle.
POWER REQUIREMENTS:
110 VAC 50-60Hz, 15 Amps or 220 VAC, 10 Amps
maximum rated for the test outlet and 0.1 Amp rated for the unit.
PHYSICAL DIMENSIONS:
Size: 8.0 X 4.5 X 2.0 inches
(20.3 X 11.4 X 5.0 cm)
Weight: 2.5 lbs (1.1 kg)
ENVIRONMENTAL:
Operating range: 59 to 95̊F (15 to 35̊
C)
Storage Temperature: 0 to 122̊
F (-
18 to 50̊C
)
Relative Humidity: 90% (max) at temperatures