Netech LKG 610 Instruction Manual Download Page 8

User Manual                                    LKG 610 

 

 

 

General Overview 

Introduction 

 

The LKG 610 is a compact Electrical Safety Analyzer designed to evaluate the electrical 

safety of all electrical equipment including medical devices and physiological 

Instrumentation.  

 

The LKG 610 measures line voltage, outlet polarity, device current, earth leakage, 

chassis resistance, leads leakage, case leakage, and performs lead ISO tests and point 

to point measurements.  

 

Each measurement is quickly and easily performed when its function key is pressed. 

The function selected and the unit of measure is indicated by an individual LED. Either 

the AAMI ESI -1993 or the IEC 601-1 test load may be selected. 

 

Resistance measurements are made using the included model 503 single conductor 

test lead.  A unique feature of the LKG 610 is its ability to verify the test readings with 

calibrated outputs of 200 

µ

 Amps and 1 Ohm.  

 

The comprehensive LKG 610 provides the best combination of features, performance, 

and value in a portable electrical safety analyzer. 

 

Key Features 

 

 

Fast, Accurate & Portable 

 

Easy to Use 

 

Large LCD Display 

 

Accessible Line Fuse 

 

Calibrated Test Points 

 

15 & 20 Amp Capability 

 

Point to Point Testing 

 

10 Lead ECG Connector 

 

Complies with both AAMI & IEC 601 Test Loads 

 

LED unit of measure Indicator 

 

 

 

 

Summary of Contents for LKG 610

Page 1: ...LKG 610 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www NetechCorp us http www PressureMeter com LKG 610 Electrical Safety Analyzer With 10 ECG Connectors INSTRUCTION MANUAL...

Page 2: ...pair and Recalibration 6 Models and Part Numbers 7 General Overview Introduction 8 Key Features 8 Specifications 9 Instrument Familiarity 10 Controls Indicators 11 Preparation for use 12 Operating Ins...

Page 3: ...otice and without incurring any obligation Netech guarantees availability of service parts for 5 years after the manufacture of the unit is discontinued The warranty is void if you elect to have the u...

Page 4: ...st procedures are found to meet those specifications when it was shipped from the factory Calibration LKG 610 is calibrated using standards traceable to National Institute of Standards and Technology...

Page 5: ...return or refund Only products returned within 60 days from the date of original purchase are eligible for refund credit In order to receive a partial refund credit of a product purchase price on a s...

Page 6: ...aging or handling Use the original carton and packaging material for shipment Returns for Refund Credit A Return Material Authorization RMA number must be obtained from our service or customer service...

Page 7: ...p www PressureMeter com Models and Part Numbers Part Number Description 535 110 10 Lead Electrical Safety Analyzer 110 V 535 220 10 Lead Electrical Safety Analyzer 220 V Standard Accessories Part Numb...

Page 8: ...ted and the unit of measure is indicated by an individual LED Either the AAMI ESI 1993 or the IEC 601 1 test load may be selected Resistance measurements are made using the included model 503 single c...

Page 9: ...SD DEVICE CURRENT 0 to 19 99 Amps Accuracy 2 Full Scale 1 LSD Line Voltage 1 to 300 Volts Accuracy 2 1 Volt TEST RECEPTACLE Hospital Grade 110 VAC 15 Amp or 220 VAC 10 Amp The rocker polarity switch s...

Page 10: ...User Manual LKG 610 10 Instrument Familiarity 2 15 1 14 12 13 11 10 9 3 4 5 6 7 8...

Page 11: ...nnection to the test receptacle 7 Polarity Switch A three position switch to select the normal and reverse polarity of the test receptacle The center position shuts the power off to the test receptacl...

Page 12: ...110 Volt devices only 2 PN 535 220 for 220 Volt devices only Before proceeding with any measurement become familiar with the measurements and the function selector switch CAUTION Make sure that the po...

Page 13: ...ice Current in Amps 3 Earth Leakage Current Measurement Press the EARTH LKG key See Figure 2 The Earth Leakage Ground Leakage Current is measured through the ground conductor of the Device Under Test...

Page 14: ...he Cord Resistance in milliohms Hammer pressure levels in excess of 20 PSI will cause damage t Figure 2 Block Diagram for Cord Resistance measurement and Case Leakage measurement external 5 Leads Leak...

Page 15: ...CAUTION During Isolation leakage measurements the LKG 610 introduces line voltage into the patient leads Even though it is current limited to 1milliAmp it is potentially lethal 7 Case Leakage Current...

Page 16: ...ASSIS When the CHS GND switch is momentarily pressed the connection will be open and the display will read zero A test lead other than the 503 may require adjustment with the zero adjustment control 3...

Page 17: ...akage current measurement the AAMI LOAD is connected in front of the INPUT AMPLIFIER and the voltage developed across it from the leakage current is measured Current Source and Amplifier Resistance is...

Page 18: ...first year warranty expires Netech will provide a second year warranty The warranty is void if the LKG 610 is serviced by anyone other than Netech or if the warranty seal is broken Calibration Servic...

Page 19: ...ssing of the service required To obtain a RMA number follow the link http www NetechCorp us and fill in the required information Email service netechcorp us or call 800 547 6557 US Canada Internationa...

Page 20: ...User Manual LKG 610 20 Appendix This section is intentionally left blank for future use...

Reviews: