Transcutaneous Pacer Test
S
e
l
e
c
t
e
d
1
0
.
0
m V
Note:
If the reading measured with the voltmeter is not
within
±
5% of the selected reading the EXPMT 2000 should
be recalibrated.
2.10.3 Calibration Check
The EXPMT 2000 can be configured as a DC voltmeter and used
to check the amplitude calibration.
A. Select Level 2 in the Main Menu.
B. Press the Aux ‘F4’key
C. Press the Cal ‘F2’ key to perform the amplitude
calibration check.
D. Select ‘Atr’ or ‘Vtr’ as the Pacer Type.
E. Input less than 2 volts into the ATR or VENT connectors
of the A.V. INPUT section on the side. The EXPMT 2000
will measure and display this voltage.
Calibration Check
Cal ‘F2’ Key Pressed
C a
l
C h e
c
k
F e
e
d
<
2
V
o l
t
s
D C
Calibration Check
C a
l
C h
e
c
k
M e
a
s
u
r
e
d
1
.
8
V
Note:
If the reading measured is not within
±
5% of the
Input voltage, the EXPMT 2000 should be recalibrated.
EXPMT 2000 User Manual
26