1. On / Off Switch
2. Serial Port
3. Five ECG Leads
4. LCD Display
5. Transvenous Inputs
6. Pacer Input Selector
7. Trancutaneous Inputs
8. Function Keys
9. DC Power Input
10. Transcutaneous Load Selector.
2.3 Power Up and Initialization
The EXPMT 2000 is battery powered. The power On / Off switch is located
on the right hand side of the instrument. When the instrument is turned on,
the model and the software revision level are displayed for about 5 seconds.
2.3.1 RS 232 Setup
The EXPMT 2000 can be connected to a printer / PC or a Palm Pilot via the
RS 232C SERIAL PORT located on the side of the instrument.
The default settings for the serial communication are:
Baud Rate = 9600, Data Bits = 8, Parity = None, Stop Bit = 1
These settings must be the same for the PC or the Printer. To retrieve the
data into a Palm Pilot use the optional interface cable for the Palm Pilot.
2.4 Menu Structure
The EXPMT 2000 is easily operated via a menu system utilizing an
eight key keypad located on the front of the instrument below the
display. Menus appear on the display above the four keys labeled ‘F1’,
‘F2’, ‘F3’, and ‘F4’. The four lower keys are labeled ‘
←
’, ‘ESC’, ‘PRN’
and ‘
→
’. Follow the instructions on the display and press the
appropriate key to use the instrument .
The ‘ESC’ key will change the display to the previous menu, and in
some situations it will confirm the menu selection. The ‘PRN’ key
activates an optional printer. The arrow keys, ‘
←
’ and ‘
→
’, move the
display forward or backward within a group of possible selections.
EXPMT 2000 User Manual
10