time the ‘

Æ

’ key is pressed the amplitude increases by 0.7 mV.  

Each time the ‘

Å

’ key is pressed the amplitude decreases by 0.7 

mV. 

 

2.5.8 Measuring Energy

 

 

A.  Press the ‘ESC’ key to return to the previous menu. 
B. Press ‘

’ key to go to Pacer Test Menu soft key Level 2 where 

energy may be selected. 

C.  Press ENR ‘F4’ key for Energy measurement. 
D.  Energy will be displayed on the top line of the display.  Energy 

will be measured continuously and updated while in the Energy 
menu. 

 

Energy Measurement Menu 

 

 E. Press the ‘ESC’ key to return to the previous (Level 2) menu. 

T  r 

n   

E  n  r 

g  =   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S  e 

n  s 

 

m m  

A  V  D   

E  n  r 

 

F1 

F2 

F3 F4 

 

 

 

 

 

 
2.6 Testing External Transvenous Pacers (AV) 

 

       The EXPMT 2000 will test both atrial and ventricular transvenous 
pacemakers.  Before performing any tests connect the pacemaker to be 
tested to either the ATR (Atrial) or VENT (Ventricular) A-V INPUT  
connectors on the top of the instrument. 

    

 

A.  Move the Pacer Input Selector switch to the A.V. INPUT position. 
B.  Turn the EXPMT 2000 on. 
C.  Select the pacer type by pressing the Atrial ‘F1’ key or the 

Ventricular ‘F2’ key. 

D.  Perform the measurements in section 2.6.1. 

 

2.6.1 Measuring Pulse Parameters 

Rate, Width, and Amplitude 

 

A.  Press the PPM Key ‘F1’ to measure the Rate. 

EXPMT 2000 User Manual

 16

 

 

Summary of Contents for EXPMT 2000

Page 1: ...EXPMT 2000 External Pacemaker Analyzer Operating Manual 110 Toledo Street Farmingdale NY 11735 Phone 631 531 0100 Fax 631 531 0101 www netech org EXPMT 2000 User Manual...

Page 2: ...etech Corporation This warranty does not apply if the product has been damaged by accident or misuse or as the result of service or modification by other than Netech Corporation or if its serial numbe...

Page 3: ...rs 13 2 5 5 Measuring Refractory Periods 13 2 5 6 Demand Sensitivity Tests 14 2 5 7 Immunity Tests 15 2 5 8 Measuring Energy 16 2 6 Testing External Transvenous Pacers AV 16 2 6 1 Measuring Pulse Para...

Page 4: ...3 3 Receiving and Saving Data 28 4 Troubleshooting EXPMT 2000 29 5 Appendix 31 This Manual is provided to explain the operation of the EXPMT 2000 External Pacemaker Analyzer It is intended for the bio...

Page 5: ...acer instruments and transvenous pacemakers being tested must be followed 4 Do not connect transthoracic and transvenous pacemakers simultaneously Test only one type at a time 5 Do not connect the pac...

Page 6: ...orage of results with optional printers and PDAs 1 1 1 Pacemaker Test Functions The EXPMT 2000 makes several measurements of external transthoracic transvenous and AV pacemakers 1 1 1 1 Transthoracic...

Page 7: ...Pulse Width 0 5 to 80 ms Accuracy 1 1 ms Refractory 20 500 ms Accuracy 2 or 10ms 1 3 1 2 Transvenous Atrial and Ventricular Test Load 500 Ohms 1 Current 1 to 25 mA Accuracy 5 5 mA Pulse Rate 30 to 800...

Page 8: ...C Adapter 110 VAC 301 110VAC AC Adapter 220 VAC 301 220 220 VAC Operating Manual 950 USER MANUAL Service Manual 950 SERV MANUAL Carrying Case 950 CASE Universal Pacer Adapters set of 2 953 Unterminate...

Page 9: ...arranty section of this manual When shipping an instrument to Netech for repair or calibration make sure that the instrument is properly packed A completed Service Return Form must be included with th...

Page 10: ...re Baud Rate 9600 Data Bits 8 Parity None Stop Bit 1 These settings must be the same for the PC or the Printer To retrieve the data into a Palm Pilot use the optional interface cable for the Palm Pilo...

Page 11: ...a n u f P R N H e a d e r Z O L L H P M D E L I F E F1 F2 F3 F4 2 5 Testing Transthoracic Pacers The EXPMT 2000 will test all transthoracic pacemakers The brand of transthoracic pacemaker to be tested...

Page 12: ...e LF 12 key to select LifePak 12 models or the Others key to select all other manufacturers E A test function is selected using F1 F2 F3 or F4 keys F Press either the or key to reach Level 2 and Level...

Page 13: ...de A Press the PPM F1 key to measure the Rate B Press the Wdth F2 key to measure the Width C Press the Amp F3 key to measure the Amplitude Note The EXPMT 2000 pacer will continuously measure the param...

Page 14: ...on is selected the measurement will start instantaneously and the STBY message is flashed until the measurement is completed Press the ESC key at any time to cancel the measurement or go back to the p...

Page 15: ...te msec msec ppm 200 300 110 400 90 100 300 140 400 110 40 25 10 250 180 300 170 350 140 400 110 2 5 7 Immunity Tests A Press the ESC key to return to the previous menu B Select Level 2 in the Main Pa...

Page 16: ...urn to the previous Level 2 menu T r n E n r g S e n s I m m A V D E n r F1 F2 F3 F4 2 6 Testing External Transvenous Pacers AV The EXPMT 2000 will test both atrial and ventricular transvenous pacemak...

Page 17: ...t Level 1 in Pacer Test Main Menu B Press the Ref L4 key for the refractory period measurement C Press the Pace F1 key for paced refractory period measurement D Press the Sense F4 key for sensed refra...

Page 18: ...F1 key to select a different waveform Sqr square Tri triangle or SSQ waveform C Press the F2 key to select a different sensitivity 100 200 10 25 and 40 msec are available D STBY will flash on the top...

Page 19: ...e immunity test Immunity Menu Imm F2 Key Pressed N o i s e I m m u n i t y 6 0 H z 5 0 H z F1 F2 F3 F4 C Press the 60Hz F1 key for a sine wave of 60Hz D Press the 50Hz F4 key for a sine wave of 50Hz E...

Page 20: ...be set to 0 1mA or higher AV Interval Menu AV D F3 Key Pressed A V D e l a y S e n s I m m A V D E n r F1 F2 F3 F4 2 6 6 Measuring Energy A Press the ESC key to return to the Level 2 Main Pacer Test...

Page 21: ...emaker by displaying the available choices The load will be changed to 50 ohms On selection of a particular pacemaker the display reverts back to the previous menu Moving the switch position back to A...

Page 22: ...ameters pacer rate pacer width pacer amplitude pacing refractory interval sensing refractory interval sensitivity and A V delay The pacer width measurement is combined with the PPM measurement For eac...

Page 23: ...Citizen or Seiko printer Seiko DPU411 type II thermal printer To print the test measurement results on the Seiko DPU 411 thermal printer follow these steps 1 Set the dip switches at the bottom of the...

Page 24: ...ons have been made press PRI in the main menu of the EXPMT 2000 and press the PC F2 key for the results to be stored in the given file name 4 The file can be printed out or edited using any text edito...

Page 25: ...within 5 of the selected reading the EXPMT 2000 should be recalibrated 2 10 2 Transcutaneous Pacer Test Function In the Transcutaneous Pacer Test Mode the EXPMT 2000 outputs a voltage of 10 millivolt...

Page 26: ...ress the Aux F4 key C Press the Cal F2 key to perform the amplitude calibration check D Select Atr or Vtr as the Pacer Type E Input less than 2 volts into the ATR or VENT connectors of the A V INPUT s...

Page 27: ...nt DST Double strike CLR Reset double strike MarR Right margin set BS Back space RST Reset to power on mode MarL Left margin set PRT Print 16 characters 2 10 5 Pacemaker Battery Current Test The EXPMT...

Page 28: ...Palm Pilot E Connect the Palm Pilot to the EXPMT 2000 using the Netech serial interface cable 650 Palm Cable F Activate the Netech PalmSoft software G Set Up 1 Baud rate 9600 2 Data bits 8 bit 3 Pari...

Page 29: ...Save the file F The file is now available for printing or transferring to a PC TROUBLESHOOTING Section 5 5 1 Troubleshooting Chart The following chart is provided for basic troubleshooting Problems o...

Page 30: ...isplay contrast is set low adapter The display shows non ASCI Characters Memory corruption Turn Unit off and then turn it On Low battery warning Battery is low Replace battery EXPMT 2000 User Manual 3...

Page 31: ...Appendix A Section 6 This Page is left blank intentionally for future updates EXPMT 2000 User Manual 31...

Reviews: