2.6.5 Measuring AV Interval
A. Press the ‘ESC’ key to return to the Level 2 Main Pacer Test
Menu.
B. Press the AV-D ‘F3’ key from the main menu for A-V Delay
measurements.
Note: The select switch must be in the A-V position and inputs
must be connected to both the ATR and VENT input jacks on
the EXPMT 2000. The amplitude values for both atrial and
ventricular output on the pacemaker must be set to 0.1mA or
higher.
AV Interval Menu: AV-D ‘F3’ Key Pressed
A V
D e
l
a
y
=
S e
n s
I
m m
A V D
E n r
F1 F2 F3
F4
2.6.6 Measuring Energy
A. Press the ‘ESC’ key to return to the Level 2 Main Pacer Test
Menu.
B. Press the Enr ‘F4’ key for Energy measurements.
Energy Measurement Menu: Enr ‘F4’ Key Pressed
A t
r
E n r
g =
S e
n s
I
m m
A V D
E n r
F1 F2 F3 F4
2.6.7 Pacemaker Type Selection
During measurements the atrial and ventricular inputs may be
changed or a different pacemaker may be tested. This is done
using the ‘Type’ function key in the Pacer Test Main Menu.
EXPMT 2000 User Manual
20