Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
13
EN
Warning
If the following defects are detected, discontinue use of the Pad:
-
Cracks or holes in the polyurethane enclosure
-
Cracked or broken Connection
-
Presence of sharp edges
Warning
Do not cut or scratch the Pad
Take care to avoid dropping or handling the Device in a way that may cause the Connection
to deform.
2. Positioning
•
Place the Pad on a flat, stable surface
3. Connection of the Pad with the Light Box
•
Connect the Pad with the Light Box by inserting the
Pad Connection (1) into the Connection Port (2).
•
Make sure the Pad and the Light Box are well
connected.
•
Make sure the cable and the light-emitting area are
not bent or stretched.
Warning
Use or manipulation of the Pad outside of these recommendations may damage the optical
fibers and affect the Pad lighting performances or lead to a risk to the patient's health.
4.C Verification of the Irradiance Value
Warning
Prior to insertion of the Pad in the disposable, measure the irradiance value on the
Pad lighting surface.
If this step is neglected, there is a risk of under-exposure
– leading to an ineffective
treatment
– or a risk of over-exposure.
Recommended Radiometer
: BiliBlanket Lightmeter II from Ohmeda (General Electrics).
For more
information on how to use the radiometer refer to its instructions for use.