Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
10
EN
1
Power
switch
2
Digital Display Screen:
-
Session Time (format: HH:MM, where HH stands for the hours and MM for
the minutes): it’s
the time remaining before the session’s end. The colon between HH and MM flashes during
an ongoing phototherapy session.
-
Total Time (format: HHHH
, which stands for the hours): it’s the total time of use of the Device.
The user can obtain this information by pressing the RESET button for 10 seconds from the
system startup.
This timer should not, under any circumstance, be used to evaluate the
phototherapy treatment duration.
3
+
and
–
controls:
to set session time. The session time can be set between 00:10 (= 10 min)
and 99:50 (= 99 h and 50 min). The minimum session time increment or decrement is 10 min.
4
RESET
button to reset session time. The user can verify that the light indicators properly light up
by pressing this button for 10 seconds from the system startup.
5
Operation indicator: blue light strip
-
Continuous light: Therapy session running
-
Flashing light: Therapy session is configured but Pad is disconnected or not properly
connected.
-
Light Off: A therapy session is neither ongoing nor configured.
6
Overheating indicator
-
Continuous light: Fan failure
-
Flashing light: Light Box overheating during treatment
7
LED failure indicator
-
Continuous light: Failure of one or more LEDs
For more information on these failures, please refer to section 5.E
“Failures and possible resolutions”.
3.C Pad Technical Features
The Pad is a woven fabric made of textile and optical fibers covered with a polyurethane enclosure. The Pad
Connection enables the transfer of blue light between the Pad and the Light Box.