1. Verification
•
Verify that the Light Box is not damaged
•
Check before each use that the power cord and the power supply do not show any abnormality.
Warning
Risk of Electric Shock
Disregarding the verification instructions may inhibit the detection of a system malfunction or
failure and lead to risks to
the patient’s health or to the user.
2. Positioning:
•
Place the Light Box on a flat, stable surface
•
Take care not to block the side (left as in the picture and right) and rear air
inlets.
•
Make sure not to position the Device or the external power supply so that
the power cord would be difficult to unplug in case of emergency.
•
Do not use adjacent to or stacked with other equipment.
•
Consider not to place the infant too close (< 30 cm
l
) to the Light Box.
Warning
If adjacent or stacked use is necessary, the Light Box should be observed to verify normal
operation in the configuration in which it will be used.
3. Connection of the Power Supply to the Light Box
•
Connect the Power Supply in the power socket
4.B Pad Setup
1. Verification and Cleaning
•
Make sure the Pad enclosure does not present sharp edges which may injure the infant.
•
Make sure the Connection is not broken
•
Clean the Pad before and after each use (see Chapter 5.C Pad Cleaning)
•
Check the number to the right of the "PAIR" symbol on both the Pad and on the Light Box labels:
these numbers must match.