Troubleshooting
NPB-40
37
Error Codes
An error code is shown when the NPB-40 detects a non
-
correctable failure.
Table 4 provides a list of error codes for the NPB-40. When one of the following
errors occurs:
•
NPB-40 sounds a low priority alarm
•
Measurements stop
•
“Er” is shown in the %SpO
2
numeric display area
•
Error code is shown in the right numeric display
•
Cycling the power should clear the shown error code
Table 4 provides a list of error codes for the NPB-40.
• Flex circuit between the front panel and
the CPU PCB is disconnected.
Inspect the flex circuit and reconnect if the flex circuit is
loose. See
Separating the Front and Back Cases
• A CPU PCB Component has failed.
Inspect the CPU PCB components and circuit board for
cracking, burning, or damage and replace the CPU PCB if
any are found. See
CPU PCB or Display PCB Replacement
• CPU PC has failed.
Replace the CPU PCB with a known good PCB. See
PCB or Display PCB Replacement
• The printer batteries are at or near a
voltage too low to operate the printer.
Install new batteries in the printer. See the printer
documentation.
• The Display PCB has failed.
Replace the Display PCB with a known good PCB.See
PCB or Display PCB Replacement
• The CPU PCB has failed.
Replace the CPU PCB with a known good PCB. See
PCB or Display PCB Replacement
• The printer is defective.
Attempt to print using a known good NPB-40. If the printer
still does not print, troubleshoot the printer. See the printer
documentation.
Table 3:
Troubleshooting Guide
(Continued)
Condition
Recommended Actione
Table 4: Error Codes
Error
Code
Scope of Action
Action
Explanation
Clinician
Re
pair
Re
tur
n
1
X
Replace SpO
2
board.
SpO
2
front end RAM error.
2
X
Replace SpO
2
board.
SpO
2
front end ROM/code integrity
error.
3
X
Replace SpO
2
board.
SpO
2
front end reported a bad CRC.
4
X
Replace SpO
2
board.
SpO
2
front end reported FSP message
not allowed.
Summary of Contents for Oximax NPB-40
Page 1: ......
Page 5: ...Contents NPB 40 iii Power Supply 81 Power Supply Monitoring 81 Index 103...
Page 8: ...Tables vi This Page Intentionally Blank...
Page 18: ...Introduction 10 This Page Intentionally Blank...
Page 22: ...Routine Maintenance 14 This Page Intentionally Blank...
Page 40: ...Performance Verification 32 This Page Intentionally Blank...
Page 50: ...Troubleshooting 42 This Page Intentionally Blank...
Page 66: ...Spare Parts 58 This Page Intentionally Blank...
Page 70: ...Packing for Shipment 62 This Page Intentionally Blank...
Page 82: ...Specifications 74 This Page Intentionally Blank...
Page 90: ...Technical Discussion 82 This Page Intentionally Blank...
Page 92: ...Technical Discussion 84 This Page Intentionally Blank...
Page 94: ...Technical Discussion 86 This Page Intentionally Blank...
Page 96: ...Technical Discussion 88 This Page Intentionally Blank...
Page 98: ...Technical Discussion 90 This Page Intentionally Blank...
Page 100: ...Technical Discussion 92 This Page Intentionally Blank...
Page 102: ...Technical Discussion 94 This Page Intentionally Blank...
Page 104: ...Technical Discussion 96 This Page Intentionally Blank...
Page 106: ...Technical Discussion 98 This Page Intentionally Blank...
Page 108: ...Technical Discussion 100 This Page Intentionally Blank...
Page 110: ...Technical Discussion 102 This Page Intentionally Blank...
Page 113: ......