P e r f o r m a n c e V e r i f i c a t i o n
Introduction
This section discusses the tests used to verify performance following repairs or
during routine maintenance. All tests can be performed without removing the
NPB-40 cover. All tests except the battery charge and battery performance tests
must be performed as the last operation before the NPB-40 is returned to the user.
If the NPB-40 fails to perform as specified in any test, repairs must be made to
correct the problem before the NPB-40 is returned to the user.
Equipment Needed
Performance Tests
The power
-
up performance test verifies that the NPB-40 is ready for patient
monitoring.
Power-On Self-Test
Discussion
Before using the NPB-40, you must verify that the NPB-40 is working properly
and is safe to use. Proper working conditions are verified each time the NPB-40 is
turned on as described in the following procedure. The verification procedure
(POST) takes 7 to 13 seconds to complete.
Caution: If any indicator or display element does not light when the NPB-40
is turned on, do not use the NPB-40. Instead, contact qualified service
personnel, your local Nellcor representative, or Nellcor’s Technical Services
Department.
Note
: Physiological conditions, medical procedures, or external agents that may
interfere with the NPB-40’s ability to detect and show measurements,
include dysfunctional hemoglobin, arterial dyes, low perfusion, dark
Equipment
Description
Digital Multimeter (DMM)
Fluke Model 87 or equivalent
Durasensor
®
O
XI
M
AX
oxygen sensor DS
-
100A
O
XI
M
AX
oxygen sensor
MAX
-
A
Safety Analyzer
Must meet current AAMI ESI/1993 & IEC
60601
-
1/1998 specifications
Extension cable
DEC-4
Stopwatch
Manual or electronic
Nellcor model SRC
-
MAX Tester
Provides testing for
O
XI
M
AX
compatible monitors
Summary of Contents for Oximax NPB-40
Page 1: ......
Page 5: ...Contents NPB 40 iii Power Supply 81 Power Supply Monitoring 81 Index 103...
Page 8: ...Tables vi This Page Intentionally Blank...
Page 18: ...Introduction 10 This Page Intentionally Blank...
Page 22: ...Routine Maintenance 14 This Page Intentionally Blank...
Page 40: ...Performance Verification 32 This Page Intentionally Blank...
Page 50: ...Troubleshooting 42 This Page Intentionally Blank...
Page 66: ...Spare Parts 58 This Page Intentionally Blank...
Page 70: ...Packing for Shipment 62 This Page Intentionally Blank...
Page 82: ...Specifications 74 This Page Intentionally Blank...
Page 90: ...Technical Discussion 82 This Page Intentionally Blank...
Page 92: ...Technical Discussion 84 This Page Intentionally Blank...
Page 94: ...Technical Discussion 86 This Page Intentionally Blank...
Page 96: ...Technical Discussion 88 This Page Intentionally Blank...
Page 98: ...Technical Discussion 90 This Page Intentionally Blank...
Page 100: ...Technical Discussion 92 This Page Intentionally Blank...
Page 102: ...Technical Discussion 94 This Page Intentionally Blank...
Page 104: ...Technical Discussion 96 This Page Intentionally Blank...
Page 106: ...Technical Discussion 98 This Page Intentionally Blank...
Page 108: ...Technical Discussion 100 This Page Intentionally Blank...
Page 110: ...Technical Discussion 102 This Page Intentionally Blank...
Page 113: ......