![Nellcor OxiMax N-65 Service Manual Download Page 46](http://html1.mh-extra.com/html/nellcor/oximax-n-65/oximax-n-65_service-manual_3917772046.webp)
Troubleshooting
40
O
XI
M
AX
N-65
Service Manual
If an error code is shown on the monitor front panel display, see Figure 3. Turn
the monitor off, wait ten seconds, then turn it back on. This clears recoverable
errors encountered by the monitor. If the error code persists, refer to Table 5:
Error Codes
, on page 43 for an explanation of the error code. Apply the action
specified.
Figure 3: Typical Error Code Display
Once you have performed the recommended action, reassemble the monitor, refer
to
Performance Verification
on page 17, and conduct a performance verification
before returning the monitor to service. If the symptom persists continue
troubleshooting.
Table 4:
Troubleshooting Guide
Condition
Recommended Action
The monitor does not turn on when you press the Power button.
• The
Power
button was not fully
depressed.
Fully press the
Power
button.
• The batteries are missing.
Open the battery compartment. If the batteries are missing,
install new batteries. Refer to the
O
XI
M
AX
N-65 Operator’s
Manual.
• The batteries are installed incorrectly.
Open the battery compartment. If the batteries are not
installed correctly, remove and reinstall them. Make sure the
batteries are not installed backwards. Refer to the
O
XI
M
AX
N-65 Operator’s Manual.
• The batteries are at, or near, a voltage too
low for the monitor to operate.
Install new batteries. Refer to the
O
XI
M
AX
N-65 Operator’s
Manual.
• The flex circuit between the front panel
and the SpO
2
PCB is damaged, or not
making a good connection.
Caution: Unlock J4 before attempting to
remove and reinsert the flex cable.
Inspect the flex cable for damage. Inspect the connection to
J4. If there is damage to the flex cable, replace the front
case. Otherwise, remove the cable from J4 and reconnect it.
Reassemble the monitor and retest. Refer to
Separating the
Front and Back Cases
on page 48 and
Front Case Assembly
Replacement
on page 54.
• The front panel keypad is defective.
Caution: Unlock J4 before attempting to
remove the flex circuit cable.
Replace the Front Case. Refer to
Front Case Assembly
Replacement
on page 54.
Summary of Contents for OxiMax N-65
Page 1: ...N 65 ...
Page 6: ...iv OXIMAX N 65 Service Manual Contents ...
Page 44: ...Performance Verification 38 OXIMAX N 65 Service Manual ...
Page 66: ...Disassembly Guide 60 OXIMAX N 65 Service Manual ...
Page 70: ...Spare Parts 64 OXIMAX N 65 Service Manual ...
Page 74: ...Packing for Shipment 68 OXIMAX N 65 Service Manual ...
Page 86: ...Specifications 80 OXIMAX N 65 Service Manual ...
Page 94: ...Blank Page ...
Page 96: ...Blank Page ...
Page 98: ...Blank Page ...
Page 100: ...Blank Page ...
Page 102: ...Blank Page ...
Page 104: ...Blank Page ...
Page 106: ...Blank Page ...
Page 108: ...Blank Page ...
Page 110: ...Blank Page ...
Page 112: ...Blank Page ...
Page 114: ...Blank Page ...
Page 117: ...Index OXIMAX N 65 Service Manual 111 Y year 13 ...
Page 118: ...Index 112 OXIMAX N 65 Service Manual ...
Page 119: ......