Performance Verification
O
XI
M
AX
N-65 Service Manual
35
2.
Perform
Test 1: BPM
on page 32.
The
Pulse Amplitude
indicator should indicate high level modulation.
3.
Perform
Test 2: SpO2
on page 33.
The
Pulse Amplitude
indicator should indicate high level modulation.
4.
Press the SRC
-
MAX %
Modulation
selection button.
The SRC
-
MAX %
Modulation
LED lights.
The monitor pulse blip bar decreases in amplitude.
The monitor:
•
displays 75 %SpO
2
•
displays 60 bpm
•
alarms
•
Pulse Amplitude
indicator displays low level modulation
5.
Perform
Test 1: BPM
on page 32.
The pulse amplitude indicator should indicate low level modulation.
6.
Perform
Test 2: SpO2
on page 33.
The pulse amplitude indicator should indicate low level modulation.
Summary of Contents for OxiMax N-65
Page 1: ...N 65 ...
Page 6: ...iv OXIMAX N 65 Service Manual Contents ...
Page 44: ...Performance Verification 38 OXIMAX N 65 Service Manual ...
Page 66: ...Disassembly Guide 60 OXIMAX N 65 Service Manual ...
Page 70: ...Spare Parts 64 OXIMAX N 65 Service Manual ...
Page 74: ...Packing for Shipment 68 OXIMAX N 65 Service Manual ...
Page 86: ...Specifications 80 OXIMAX N 65 Service Manual ...
Page 94: ...Blank Page ...
Page 96: ...Blank Page ...
Page 98: ...Blank Page ...
Page 100: ...Blank Page ...
Page 102: ...Blank Page ...
Page 104: ...Blank Page ...
Page 106: ...Blank Page ...
Page 108: ...Blank Page ...
Page 110: ...Blank Page ...
Page 112: ...Blank Page ...
Page 114: ...Blank Page ...
Page 117: ...Index OXIMAX N 65 Service Manual 111 Y year 13 ...
Page 118: ...Index 112 OXIMAX N 65 Service Manual ...
Page 119: ......