Performance Verification
O
XI
M
AX
N-65 Service Manual
31
Test Setup
You must complete Test Setup before performing any other pulse oximetry
functional tests.
To perform test setup:
1.
Press the
Power
button to turn on the monitor.
2.
After the monitor presents the monitoring screen, connect the SRC
-
MAX
tester to the monitor sensor port as shown.
Note:
Use a DEC-4 extension cable between the SRC-MAX and the
monitor.
The SRC-MAX panel lights and in a few seconds the monitor tracks the
simulator.
After the monitor completes POST, the monitor:
•
is in SpO
2
alarm
•
displays an %SpO
2
of 75 (pass criteria is 73 to 77 %SpO
2
inclusive)
•
displays a pulse rate of 60 (pass criteria is 57 to 63 bpm inclusive)
1 — Sensor Port
Summary of Contents for OxiMax N-65
Page 1: ...N 65 ...
Page 6: ...iv OXIMAX N 65 Service Manual Contents ...
Page 44: ...Performance Verification 38 OXIMAX N 65 Service Manual ...
Page 66: ...Disassembly Guide 60 OXIMAX N 65 Service Manual ...
Page 70: ...Spare Parts 64 OXIMAX N 65 Service Manual ...
Page 74: ...Packing for Shipment 68 OXIMAX N 65 Service Manual ...
Page 86: ...Specifications 80 OXIMAX N 65 Service Manual ...
Page 94: ...Blank Page ...
Page 96: ...Blank Page ...
Page 98: ...Blank Page ...
Page 100: ...Blank Page ...
Page 102: ...Blank Page ...
Page 104: ...Blank Page ...
Page 106: ...Blank Page ...
Page 108: ...Blank Page ...
Page 110: ...Blank Page ...
Page 112: ...Blank Page ...
Page 114: ...Blank Page ...
Page 117: ...Index OXIMAX N 65 Service Manual 111 Y year 13 ...
Page 118: ...Index 112 OXIMAX N 65 Service Manual ...
Page 119: ......