Never use equipment that has parts missing or equipment that might contain loose parts
inside of it (that is, inside an enclosed portion of the equipment). If you suspect a piece of
equipment has missing or loose parts, contact Natus. Routinely inspect system cables and
components for regular wear and tear.
Perform the recommended maintenance. Refer to the
Maintenance & Cleaning
section for
further details.
Do not immerse the amplifier or any of its components in water or other fluid.
This equipment/system is intended for use by Healthcare professionals ONLY. Please read
this section before installing any of the hardware. Refer to this section when you operate,
transport, store, or re-install the system.
Proper use of this device depends on the careful reading of all instructions and labels that
come with or on the system. Inaccurate measurements may be caused by incorrect
application or use.
When replacing the fuse for the Quantum amplifier system, it must be replaced by a fuse
with the same type and rating as the original fuse. Replacement fuses should be
The Quantum Amplifier system is compatible with NeuroWorks/SleepWorks v9.0 and
newer.
The Natus base unit is classified as an IPX0
– ordinary degree of protection against ingress
of water according to IEC 60529.
The Quantum amplifier Breakout is classified as body worn and has an IPX1 rating while
inside the pouch (-61) and meets IEC 60601-2-26 spillage requirement without the pouch.
The Quantum amplifier system is classified as a class I device according to IEC 60601-1.
WARNING:
Third-party software installed on the acquisition computer may interfere with
the operation of the Natus software. Please consult Natus Technical Support before
installing third-party software on the computer.
WARNING:
No modification of this equipment is allowed.
For systems used in the Operating Room with a Cautery device, the use of Ethernet type
connection is mandatory to ensure continuous recording.