6-4
OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
Repair Procedures
Reassembling the Smart Scale
After performing a service procedure, reassemble the Smart Scale and verify that it
operates correctly before use.
Required Items:
Allen wrench
To reassemble to scale:
1
Attach the weighing platform:
a
Place the spacer on the base (see Figure 6.3).
b
Place the weighing platform on the base, then secure with the two
socket-head cap screws and the two flat-head cap screws.
Apply 90-in.–lb. torque to the two socket-head cap screws.
Tighten the two flat-head cap screws firmly; these screws require
minimal force to secure.
c
Slide one end of the scale so that it hangs over the edge of the table, then
secure the four socket-head cap screws underneath the device.
Apply 5-in.–lb. ±1 torque to the cap screws that attach the main platform
to the chassis.
d
Repeat step c on the other end.
2
Calibrate the scale (see page 4-3).
NOTE
#
When reinstalling the weighing platform, confirm that the channel connected to the right
load cell is square and that the swing arms move freely.
CAUTION
Always recalibrate the Smart Scale after disassembly/reassembly (see
Summary of Contents for OLYMPIC SMART SCALE 50
Page 2: ......
Page 4: ......
Page 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...
Page 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...
Page 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...
Page 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...
Page 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...
Page 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...
Page 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...