OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
4-3
Calibrating the Smart Scale
Calibrating the Smart Scale
As with any precision electronic instrument, the scale can drift out of calibration
over time. Calibrate the Smart Scale before first use and every six months.
Required Items:
Calibration weights (part no. 56357)
To calibrate the Smart Scale:
1
Place the scale on a firm, level surface and protect it from moving air. Confirm
that the ambient temperature is 70–75°F (21.1–23.9°C).
2
Press and hold
ZERO
, then press
ON/OFF
to enter calibration mode
(see Figure 4.1); confirm the following:
The current software version (e.g.,
CAL 1.0
) appears.
C 0
appears while the scale captures the zero point.
C 12500
appears prompting for placement of the 12.5 kg calibration weight.
3
Place the two calibration weights (totaling 12,500 grams) on the center of the
platform. Calibration is successfully completed when
12.50
kg displays (see
Figure 4.1
Calibration mode displays
4
Remove the calibration weight from the scale, then press
ZERO
.
5
Press
ON/OFF
to turn off the scale.
1) Displaying the software version
2) Capturing the zero point
3) Requesting the calibration weight
4) Displaying the weight of the
calibration weight
Summary of Contents for OLYMPIC SMART SCALE 50
Page 2: ......
Page 4: ......
Page 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...
Page 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...
Page 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...
Page 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...
Page 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...
Page 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...
Page 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...