Error code E905
(Check Sum Error)
Non-volatile memory
corrupted or damaged
Calibrate the scale (see
Test the main PCB (see
replace if necessary (see page 6-12).
Unreadable
display/garbled
Electrical fast transient
bursts
Install a filter on power input lines.
Contact Natus Medical (see page 5-4).
Table 5.1
Troubleshooting chart, continued
Problem
Probable Cause
Solution