Madsen AccuScreen OAE & ABR Screener User Manual
86
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems within Professional Healthcare use environment
MADSEN AccuScreen is intended for use in the electromagnetic environment specified below. The user of MADSEN AccuScreen should ensure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
3 V rms
150 kHz to 80 MHz
6 V rms
IISM Bands
6 V rms
IISM Bands
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Proximity fields from RF wire-
less communications
IEC 61000-4-3
27 V/m
385 MHz
28 V/m
450 MHz,
27 V/m
385 MHz
28 V/m
450 MHz,
Separation distance between any electronic parts of
MADSEN AccuScreen and any RF wireless
communication equipment must be more than 30 cm
(11.8 inches).
9 V/m
710 MHz, 745 MHz, 780 MHz
9 V/m
710 MHz, 745 MHz, 780 MHz
Note
: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
28 V/m
810 MHz, 870 MHz, 930 MHz,
28 V/m
810 MHz, 870 MHz, 930 MHz,
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
1720 MHz, 1845 MHz, 1970 MHz
28 V/m
2450 MHz,
28 V/m
2450 MHz,
9 V/m
5240 MHz, 5500 MHz, 5785 MHz
9 V/m
5240 MHz, 5500 MHz, 5785 MHz