Madsen AccuScreen OAE & ABR Screener User Manual
84
3.9
Notes on EMC (Electromagnetic Compatibility)
•
MADSEN AccuScreen is part of a medical electrical system and is thus subject to special safety precautions. For this
reason, the installation and operating instructions provided in this document should be followed closely.
•
Portable and mobile high-frequency communication devices, such as mobile phones, may interfere with the func-
tioning of MADSEN AccuScreen.
Warning: Follow guidance and manufacturer's declaration – electromagnetic emissions for all equipment and systems.
Caution: MADSEN AccuScreen is intended for use in the electromagnetic environment specified below. The user of
MADSEN AccuScreen should ensure that it is used in such an environment.
IEC 60601-1-2:2014 and EN 60601-1-2:2015
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
MADSEN AccuScreen is intended for use in the electromagnetic environment specified below. The user of MADSEN AccuScreen should ensure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR11
Group 1
MADSEN AccuScreen uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR11
Class B
MADSEN AccuScreen is suitable for use in all environments, including domestic environments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Not applicable